静态导板牙髓治疗精度影响因素分析

Analysis of Factors Affecting the Accuracy of Static-guided Endodontics.
PubMed 文献 · J Endod · 2025-10-25

本研究评估模板类型、套管高度、转速、牙位、钻孔距离和钻头直径对静态导板牙髓治疗精度的影响。体外实验使用113颗离体前磨牙和磨牙及10颗3D打印树脂牙,通过CBCT和光学扫描共注册测量偏差。结果显示树脂模板(无金属套管)角度偏差更大,套管高度影响根尖线性偏差,钻孔距离增加根尖偏差增大,而钻头直径、转速和牙位不影响精度。

要点

  • 树脂模板(无金属套管)角度偏差大于金属模板,根尖线性偏差大于带金属套管树脂模板。
  • 4mm高套管树脂模板根尖线性偏差大于6mm和8mm高套管,6mm与8mm间无显著差异。
  • 根尖线性偏差随钻孔距离增加而增大,钻头直径、转速和牙位不影响精度。

核心结论

  • 根尖线性偏差随钻孔距离增加而增大。
  • 不推荐使用无金属套管的树脂模板。
  • 静态导板牙髓治疗精度不受钻头直径、转速或牙位(树脂或离体牙)影响。

研究设计与样本

  • 体外研究,使用113颗离体前磨牙和磨牙及10颗3D打印树脂牙,安装在标准牙模对应牙位。
  • 获取锥形束CT和光学扫描并共注册,设计导板并使用Implant Studio测量偏差。

主要发现

  • 树脂模板(无金属套管)角度偏差高于金属模板(P < .05)。
  • 树脂模板(无金属套管)根尖线性偏差显著高于树脂模板(带金属套管)(P < .05)。
  • 金属模板冠部线性偏差高于树脂模板(带金属套管)(P < .05)。
  • 4mm高套管树脂模板根尖线性偏差高于6mm和8mm高套管(P < .05),6mm与8mm间无显著差异。
  • 钻头直径、转速和牙位(树脂或离体牙)不影响静态导板牙髓治疗精度。

临床怎么用

  • 临床应避免使用无金属套管的树脂模板。
  • 选择套管高度时,6mm或8mm可能优于4mm以减少根尖偏差。
  • 钻孔距离增加时需注意根尖线性偏差可能增大。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: To evaluate how template type, sleeve height, revolutions per minute (RPM), tooth type, drilling distance, and drill diameter affect the accuracy of static-guided endodontics. METHODS: In this in vitro study, 113 isolated premolars and molars collected from the oral surgery clinic and 10 three-dimensional printed resin teeth were mounted in a standard dental mold in their corresponding tooth positions. Cone-beam computed tomography and optical scans were obtained and co-registered. Templates were designed, and deviations were measured using Implant Studio (3Shape Unite, 3shape, Copenhagen, Denmark). RESULTS: The angular deviation of the resin template (without metal sleeve) was higher than that of the metal template (P < .05). The apical linear deviation of the resin template (without metal sleeve) was significantly higher than that of the resin template (with metal sleeve) (P < .05). The coronal linear deviation of the metal template was higher than that of the resin template (with metal sleeve) (P < .05).The apical linear deviation of the resin template with 4-mm height sleeve was higher than that of the 6-mm and 8-mm height sleeves (P < .05), and there was no significant difference between the 6-mm and 8-mm height sleeves. Drill diameter, RPM, and tooth type (resin or isolated teeth) did not affect the accuracy of static guided endodontics. CONCLUSIONS: The apical linear deviation increased with drilling distance. Resin templates (without metal sleeves) are not recommended. The accuracy of static-guided endodontics was not affected by drill diameter, RPM, or tooth type (resin or isolated teeth).