解剖参数对C形根管电子根尖定位仪准确性的影响:结合Feret直径的微CT新分析
本研究用显微CT结合Feret直径分析42颗下颌C形磨牙的119个根管,评估Raypex 6和Root ZX II定位根尖孔与根尖狭窄的准确性。两种设备定位根尖孔±0.5 mm的准确率分别为97.5%和96.4%,但定位根尖狭窄的准确性显著较低,多根尖狭窄形态是唯一负影响因素。
要点
- Raypex 6和Root ZX II在C形根管中定位根尖孔准确,但定位根尖狭窄可靠性较低。
- 多根尖狭窄形态是降低两种设备根尖狭窄定位准确性的唯一因素。
- 电子根尖定位仪可能探测根管最靠冠方的狭窄处而非最窄处。
核心结论
- Raypex 6和Root ZX II在下颌C形根管磨牙中能准确检测根尖孔,但对根尖狭窄的检测可靠性较低。
- 根尖狭窄准确性降低与多根尖狭窄形态相关,提示电子根尖定位仪可能检测根管根尖方向第一个狭窄部而非最窄部。
研究设计与样本
- 体外研究,采用视觉法测量实际根管长度,用Raypex 6和Root ZX II获取电子根管长度(确定根尖孔)和电子工作长度(确定根尖狭窄)。
- 以根工作长度(根管长度减去AC-AF距离)为参考评估电子工作长度准确性,解剖参数来自显微CT扫描。
- 纳入42颗C形磨牙的119个根管,统计分析采用Fisher精确检验和多因素logistic回归。
主要发现
- Raypex 6和Root ZX II确定根尖孔在±0.5 mm内的根管比例分别为97.5%和96.4%。
- 检测根尖孔与根尖狭窄的准确性之间差异有统计学意义(P < .05)。
- 多因素logistic回归后,多根尖狭窄形态是唯一对两种电子根尖定位仪根尖狭窄准确性产生负影响的因素。
临床怎么用
- 在C形根管下颌磨牙中,电子根尖定位仪可可靠定位根尖孔,但确定根尖狭窄时需谨慎。
- 遇多根尖狭窄形态时,电子根尖定位仪的工作长度准确性可能降低,应结合其他方法确认。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: This study aimed to evaluate the influence of anatomical parameters on the accuracy of 2 electronic apex locators (Raypex 6 and Root ZX II) in C-shaped canal mandibular molars using a novel micro-computed tomography (micro-CT) analysis incorporating Feret diameter. METHODS: The actual root canal length was measured using the visual method. The electronic canal length, which determined the apical foramen (AF), and the electronic working length, which determined the apical constriction (AC), were obtained using Raypex 6 and Root ZX II. The root working length, calculated by subtracting the AC-AF distance from the root canal length, served as a reference to evaluate the accuracy of the electronic working length. Anatomical parameters were derived from micro-CT scans and included Feret diameters of AC and AF, AC-AF distance, presence/absence of accessory canal, C-shaped type, AF position and AC morphology. Statistical analysis involved Fisher's exact test and multinomial logistic regression. RESULTS: The study included 42 C-shaped molar teeth and 119 canals. Raypex 6 and Root ZX II determined AF within ±0.5 mm in 97.5% and 96.4% of canals, respectively. A statistically significant difference was found in accuracy between detecting AF and AC (P < .05). After multinomial logistic regression, multiple AC morphology was the only factor negatively affecting AC accuracy in both electronic apex locators. CONCLUSIONS: Raypex 6 and Root ZX II accurately detected AF but were less reliable for AC in C-shaped canal mandibular molars. Reduced AC accuracy was associated with multiple AC morphology, indicating that electronic apex locators may detect the first narrow part of the canal apically rather than the narrowest part.
引用
- Tung KW, Sasaki H, Cavalcanti B, et al. Influence of Anatomical Parameters on the Accuracy of Electronic Apex Locators in C-shaped Canals: A Novel Micro-CT Analysis Incorporating Feret Diameter. J Endod. 2025. PMID: 41043606. doi:10.1016/j.joen.2025.09.012
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。