文献速递口腔修复与种植

全烧结氧化锆混合基台冠在Ti-base上的拉出力与断裂载荷:表面处理与材料比较

Pull-out force and fracture load of fully sintered zirconia hybrid abutment-crowns on Ti-base abutments: Surface treatment effects and material comparison.
PubMed 文献 · J Dent · 2025-09-25

本研究评估表面处理对含氧化铌钇稳定四方氧化锆多晶([Y, Nb]-TZP)全烧结混合基台冠拉出力的影响,并与二硅酸锂(LD)和聚合物渗透陶瓷网络(PICN)比较断裂载荷。结果显示含10-MDP底涂剂可提高拉出力,且该材料断裂载荷与LD和PICN相当。

要点

  • 含10-MDP底涂剂可显著提高全烧结(Y, Nb)-TZP混合基台冠的拉出力。
  • 仅喷砂处理组拉出力最低,提示单独喷砂可能不利于 retention。
  • 全烧结(Y, Nb)-TZP断裂载荷与LD和PICN无显著差异。

核心结论

  • 含10-MDP底涂剂可提高全烧结(Y, Nb)-TZP混合基台冠的拉出力,无论是否联合喷砂。
  • 该材料制作的Ti-base支持种植修复体断裂载荷与LD和PICN修复体相当。

研究设计与样本

  • 拉出力实验:40个(Y, Nb)-TZP混合基台冠试件,按表面处理分4组(n=10):无处理、喷砂、10-MDP底涂剂、喷砂+10-MDP底涂剂。
  • 断裂载荷实验:30个混合基台冠试件,分别由(Y, Nb)-TZP、LD和PICN块研磨而成(n=10)。
  • 所有试件粘接后经热循环,分别测试拉出力或断裂载荷并记录失败模式。

主要发现

  • FSZ-M和FSZ-AM组拉出力显著高于FSZ-N组。
  • FSZ-A组拉出力最低。
  • LD组断裂载荷高于PICN组。
  • FSZ组断裂载荷与LD或PICN组相比无显著差异。

关键数值与推荐

  • 拉出力比较:FSZ-M和FSZ-AM显著高于FSZ-N;FSZ-A最低。
  • 断裂载荷比较:LD高于PICN;FSZ与LD、PICN无显著差异。

临床怎么用

  • 对全烧结氧化锆混合基台冠粘接前使用含10-MDP底涂剂以增强固位。
  • 该材料断裂载荷与LD和PICN相当,支持其用于Ti-base基台的潜在临床可行性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: To evaluate the effect of surface treatment on the pull-out force of implant restorations supported by Ti-bases and milled from fully sintered niobium oxide-containing yttria-stabilized tetragonal zirconia polycrystals ([Y, Nb]-TZP), and to compare their fracture load with that of lithium disilicate (LD) and polymer-infiltrated ceramic network (PICN) restorations. METHODS: To evaluate pull-out force, 40 (Y, Nb)-TZP hybrid abutment-crown specimens were divided into four groups based on the surface treatment applied (n = 10): no surface treatment (FSZ-N group), airborne-particle abrasion (APA) (FSZ-A group), a 10-methacryloyloxydecyl dihydrogen phosphate (MDP) primer (FSZ-M group), and both APA and a 10-MDP primer (FSZ-AM group). After cementation and thermocycling, the pull-out forces and failure modes were examined by pull-out testing. To evaluate fracture load, 30 hybrid abutment-crown specimens were milled from (Y, Nb)-TZP (fully sintered zirconia; FSZ group), LD (LD group), and PICN (PICN group) blocks (n = 10). Following cementation and thermocycling, fracture loads and failure modes were recorded. RESULTS: The FSZ-M and FSZ-AM groups demonstrated significantly higher pull-out forces than the FSZ-N group. The FSZ-A group showed the lowest pull-out force. The LD group had a higher fracture load than the PICN group. The fracture load of the FSZ group showed no significant difference compared with the LD or PICN group. CONCLUSIONS: Applying a 10-MDP-containing primer enhanced the pull-out force of hybrid abutment-crowns milled from fully sintered (Y, Nb)-TZP, with or without APA. Implant restorations fabricated from this material and supported by Ti-base abutments exhibited a fracture load comparable to that of LD and PICN restorations. CLINICAL SIGNIFICANCE: For implant restorations milled from fully sintered zirconia and supported by Ti-base abutments, applying a 10-MDP-containing primer is a key step for improving retention. Furthermore, their fracture load, which is comparable to other restorations milled from established materials, supports their potential for clinical use on Ti-base abutments.

来源:J Dent,查看原文。版权归原作者所有。

引用

  1. Wan Q, Lee DH, Kim RJY, et al. Pull-out force and fracture load of fully sintered zirconia hybrid abutment-crowns on Ti-base abutments: Surface treatment effects and material comparison. J Dent. 2025. PMID: 41015219. doi:10.1016/j.jdent.2025.106134

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。