智能手机脉搏血氧仪用于牙髓活力客观评估的临床验证:一项诊断准确性研究
结果显示两种设备均能显著区分牙髓炎症阶段及坏死,PO2诊断性能与PO1相当,且具有数据采集效率和临床工作流程整合优势。
要点
- PO1与PO2测量的%SpO2在可复性牙髓炎、不可复性牙髓炎和牙髓坏死组间差异显著(P < .05)。
- PO1在所有临床条件下记录的氧饱和度均高于PO2(P < .05), discriminatory capability更优。
- 智能手机适配血氧仪诊断性能与PO1相当,且便于数据采集与工作流程整合。
核心结论
- 两种脉搏血氧仪对牙髓活力评估及牙髓炎症阶段和坏死的区分均具有显著临床价值。
- 智能手机适配血氧仪在诊断性能上与PO1相当,并提供数据采集效率和临床工作流程整合优势。
研究设计与样本
- 每颗患牙另测2颗健康牙(邻牙和对侧牙)作为健康对照组,共比较PO1和PO2两种脉搏血氧仪。
- 统计分析采用单因素方差分析及Tukey事后检验。
主要发现
- PO1测量的平均%SpO2:可复性牙髓炎87.7%,不可复性牙髓炎80.5%,牙髓坏死73.2%,健康牙91.89%。
- PO2测量的平均%SpO2:可复性牙髓炎85.6%,不可复性牙髓炎79.3%,牙髓坏死72.1%,健康牙90.1%。
- 可复性牙髓炎、不可复性牙髓炎和牙髓坏死组间%SpO2差异显著(P < .05),健康对照亚组(邻牙与对侧牙)间无显著差异。
- PO1在所有临床条件下记录的氧饱和度均高于PO2(P < .05), discriminatory capability更优。
关键数值与推荐
- PO1平均%SpO2:可复性牙髓炎87.7%、不可复性牙髓炎80.5%、牙髓坏死73.2%、健康牙91.89%。
- PO2平均%SpO2:可复性牙髓炎85.6%、不可复性牙髓炎79.3%、牙髓坏死72.1%、健康牙90.1%。
- 组间比较采用单因素方差分析及Tukey事后检验,显著性水平P < .05。
临床怎么用
- 智能手机连接脉搏血氧仪可作为牙髓活力客观评估的便捷替代工具,辅助区分牙髓炎症阶段与坏死。
- 临床使用时需注意PO2读数系统性低于PO1,判读时应参照设备对应的参考范围。
局限性
- 原文未明确报告样本量计算、盲法设计及设备测量重复性等细节。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Pulse oximetry exhibits great potential for use in endodontic diagnosis as an effective method to assess pulp vitality. Cell phone-integrated oximeters represent an emerging alternative that may offer greater accessibility. This study aimed to investigate the relation between pulp oxygenation rates (%SpO2) and clinical diagnosis of healthy pulp (HP), reversible pulpitis (RP), symptomatic irreversible pulpitis (IP), or pulp necrosis (PN), comparing 2 pulse oximeters (conventional [PO1] and mobile-connected). METHODS: Thirty affected premolars were grouped according to the clinical diagnosis of RP, IP, or PN and compared to vitality diagnosis by %SpO2 obtained by both pulse oximeters: PO1 and by smartphone-based (PO2). For each affected tooth, 2 additional teeth (adjacent and contralateral) with HP were tested. Statistical analysis was performed using one-way ANOVA with Tukey's post-hoc test. RESULTS: The mean %SpO2 values measured by PO1 were 87.7% for RP, 80.5% for IP, 73.2% for PN, and 91.89% for HP. Corresponding values obtained with PO2 were 85.6% (RP), 79.3% (IP), 72.1% (PN), and 90.1% (HP). Significant differences in %SpO2 were observed among RP, IP, and PN groups (P < .05), but not between healthy control subgroups (adjacent and contralateral). PO1 consistently recorded higher oxygen saturation than PO2 across all clinical conditions (P < .05), demonstrating superior discriminatory capability. CONCLUSIONS: Both pulse oximeters demonstrated significant clinical value for pulp vitality assessment and differentiation of pulpal inflammation stages and necrosis. The smartphone-adapted oximeter has proven to be comparable to PO1 in diagnostic performance, while offering data collection efficiency and clinical workflow integration.