电抛光对旋转与往复镍钛器械力学及冶金性能的体外评价

Mechanical and Metallurgical In Vitro Evaluation of Electropolished Versus Non-Electropolished Rotary and Reciprocating Instruments.
PubMed 文献 · Int Endod J · 2025-08-04

本研究比较电抛光与非电抛光的One RECI(往复)和One Curve mini(旋转)镍钛器械,二者几何与热处理相同。结果显示电抛光提升两类器械柔韧性,但仅降低往复器械抗扭强度,表面微硬度和切割效率无变化。

要点

  • 电抛光可提升两类镍钛器械的柔韧性,降低弯曲与屈曲载荷。
  • 电抛光仅降低往复One RECI的抗扭强度,旋转器械无显著差异。
  • 电抛光不影响两类器械的表面微硬度和切割效率。

核心结论

  • 电抛光改善One RECI和One Curve mini镍钛器械的柔韧性,不损害表面微硬度和切割效率。
  • 电抛光对抗扭强度的影响具有系统依赖性,仅往复One RECI器械抗扭强度降低。

研究设计与样本

  • 体外研究比较电抛光与非电抛光的One RECI(往复)和One Curve mini(旋转)镍钛器械,几何与热处理相同。
  • 采用光镜和扫描电镜分析设计,EDS和DSC进行冶金表征,力学测试包括扭转、弯曲、屈曲、微硬度和切割效率。
  • 统计采用独立样本t检验或Mann-Whitney U检验,显著性设为p < 0.05。

主要发现

  • 设计与冶金分析确认各组内电抛光与非电抛光器械在几何、横截面、尖端、元素组成和相变温度上等效。
  • 电抛光显著降低两类器械的弯曲载荷和屈曲阻力(p < 0.05),提示柔韧性增强。
  • 电抛光One RECI器械的最大扭矩和断裂前旋转角度显著降低(p < 0.05)。
  • One Curve mini组扭转性能无显著差异(p > 0.05)。
  • 两类器械表面微硬度和切割效率在电抛光前后无显著差异(p > 0.05)。

临床怎么用

  • 电抛光镍钛器械可能提供更好的柔韧性,适用于弯曲根管。
  • 使用电抛光往复One RECI器械时需注意其抗扭强度降低,可能影响抗折断性能。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIM: To evaluate the effect of electropolishing on the mechanical properties of One RECI and One Curve mini nickel-titanium (NiTi) instruments by comparing electropolished and non-electropolished versions of each instrument type. METHODOLOGY: Electropolished and non-electropolished One RECI (reciprocating) and One Curve mini (rotary) NiTi instruments, all manufactured with identical geometry and heat treatment, were evaluated. Instrument design was analysed by light microscopy and scanning electron microscopy, while metallurgical characterisation was performed using energy-dispersive X-ray spectroscopy (EDS) and differential scanning calorimetry (DSC). Mechanical performance was assessed through torsional resistance, bending and buckling load, surface microhardness, and cutting efficiency. Statistical comparisons were performed using the independent samples t-test or the Mann-Whitney U-test, with significance set at p < 0.05. RESULTS: Design and metallurgical analyses confirmed that electropolished and non-electropolished instruments within each group were equivalent in terms of geometry, cross-sectional design, tip configuration, elemental composition, and phase transformation temperatures. Electropolishing significantly enhanced flexibility in both instrument types, as indicated by reduced bending loads and lower buckling resistance (p < 0.05). However, torsional strength was significantly reduced in the electropolished One RECI instruments, reflected by lower maximum torque and angle of rotation prior to fracture (p < 0.05). No significant torsional differences were observed in the One Curve mini group (p > 0.05). Surface microhardness and cutting efficiency remained unaffected by electropolishing in both systems (p > 0.05). CONCLUSIONS: Electropolishing improved the flexibility of both One RECI and One Curve mini NiTi instruments without compromising their surface microhardness or cutting efficiency. However, its impact on torsional resistance was system-dependent, resulting in reduced strength only in the reciprocating One RECI instruments.