基于电阻法的镍钛旋转器械循环疲劳评估

Cyclic Fatigue Evaluation of Nickel-Titanium Rotary Instruments Using an Electrical Resistance-based Method.
PubMed 文献 · J Endod · 2025-06-26

这项体外研究探讨电阻与牙髓镍钛旋转器械循环疲劳的相关性。选用几何相同的ProTaper Universal F2和ProTaper Gold F2,在人工根管中旋转至断裂记录NCF,并在20%、40%、60%、80% NCF后高压灭菌,每10次旋转及灭菌前后测电阻。结果显示电阻随疲劳程度呈二次方增加,80% NCF时PTU F2和PTG F2电阻分别增加2.38%和2.63%,高压灭菌无显著影响。

要点

  • 电阻随镍钛器械循环疲劳程度呈二次方增加。
  • 80% NCF时PTU F2和PTG F2电阻分别增加2.38%和2.63%。
  • 高压灭菌对电阻无显著影响。

核心结论

  • 电阻可作为量化指标评估和预测镍钛旋转器械的循环疲劳程度,且似乎不受高压灭菌影响。

研究设计与样本

  • 体外研究,选用几何相同的ProTaper Universal F2和ProTaper Gold F2,分循环疲劳测试组和灭菌测试组。
  • 循环疲劳测试组在人工根管中旋转至断裂,记录断裂周期数(NCF)。
  • 灭菌测试组旋转至20%、40%、60%、80% NCF后高压灭菌,每10次旋转及灭菌前后测电阻,并用扫描电镜分析表面微裂纹。

主要发现

  • 电阻随循环疲劳程度呈二次方增加,伴随微裂纹扩展。
  • 80% NCF时,PTU F2电阻增加2.38%,PTG F2增加2.63%。
  • 高压灭菌对电阻无显著影响。

关键数值与推荐

  • 80% NCF时电阻增加:PTU F2为2.38%,PTG F2为2.63%。
  • 电阻测量频率:每10次旋转及灭菌前后。

临床怎么用

  • 电阻法或可用于临床量化评估镍钛器械循环疲劳程度。
  • 高压灭菌后电阻测量值可能仍具参考价值。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: This in vitro study aimed to establish the correlation between electrical resistance and cyclic fatigue of endodontic nickel-titanium (NiTi) rotary instruments as a quantifiable cyclic fatigue indicator. METHODS: Two kinds of geometrically identical NiTi rotary instruments ProTaper Universal F2 and ProTaper Gold F2 were selected and allocated into 2 experimental groups: a cyclic fatigue test group and a sterilization test group. In the cyclic fatigue test group, instruments were rotated in an artificial root canal until fracture, with the number of cycles (rotations) to fracture (NCF) recorded. In the sterilization test group, instruments were rotated to 20%, 40%, 60%, and 80% of the NCF, followed by autoclaving sterilization. Electrical resistance was measured after each 10 rotations, as well as pre- and post-sterilization. Scanning electron microscopy was applied to analyze the instrument surfaces, focusing on the expanding of microcracks over rotations and the effect of autoclaving. The collected data were then statistically analyzed. RESULTS: The results indicated that the electrical resistance of tested instruments increased quadratically with the cyclic fatigue degrees along with the expanding of microcracks. At the 80% NCF, the electrical resistance increased by 2.38% and 2.63% for PTU F2 and PTG F2 respectively. Autoclaving had no significant effect on electrical resistance. CONCLUSION: Electrical resistance could be used as a quantifiable indicator to evaluate and predict the cyclic fatigue degree of NiTi rotary instruments and seems unaffected by autoclaving.