聚合物、钢制与钛制超声工作尖在根管内药物去除中的频率与性能分析

Frequency and Performance Analysis of Polymer, Steel, and Titanium Ultrasonic Inserts in Intracanal Medication Removal.
PubMed 文献 · J Endod · 2025-06-16

本研究用声波分析评估钢制、钛制与聚合物超声工作尖产生的频率,并用显微CT比较iVac系统、被动超声冲洗(PUI)与常规冲洗去除氢氧化钙基根管内药物的效果。结果显示各工作尖频率在37至45 kHz之间;iVac组药物去除率99.8 ± 0.3%,高于PUI组97.3 ± 2.9%,两者均显著优于常规冲洗73.0 ± 9.4%(P < .05)。

要点

  • 钢制、钛制与聚合物超声工作尖产生的频率在37至45 kHz之间。
  • iVac组药物去除率99.8 ± 0.3%,高于PUI组97.3 ± 2.9%。
  • iVac与PUI均显著优于常规冲洗73.0 ± 9.4%(P < .05)。

核心结论

  • 各超声工作尖的超声功率处于预期频率范围内,iVac系统在去除根管内药物方面表现出更优效果。

研究设计与样本

  • 体外研究,对Ultra X系统(蓝色、银色、金色)的钢制与钛制工作尖及iVac系统的聚合物工作尖进行振动声学分析。
  • 在Ultra X(45 kHz)与Woodpecker(40 kHz)超声设备中,工作尖置于水槽内标准化定位,用MATLAB软件生成激励信号,功率测量重复两次。
  • 30颗牙根管预备后封入氢氧化钙基药物并显微CT扫描,7天后先用K锉(45.02号)初步去除,再分常规冲洗、PUI或iVac三组处理,随后第二次显微CT扫描。
  • 采用方差分析与Tukey检验比较超声功率与残留药物体积。

主要发现

  • 钢制、钛制与聚合物工作尖产生的频率为37至45 kHz。
  • iVac组药物去除率为99.8 ± 0.3%,PUI组为97.3 ± 2.9%,常规冲洗组为73.0 ± 9.4%。
  • iVac组与PUI组均显著优于常规冲洗组(P < .05)。

关键数值与推荐

  • iVac组药物去除率99.8 ± 0.3% vs PUI组97.3 ± 2.9% vs 常规冲洗组73.0 ± 9.4%(P < .05)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: Ultrasonic inserts can be manufactured in various shapes, angles, and materials, this study evaluated the frequencies generated by steel, titanium, and polymer ultrasonic inserts using acoustic wave analysis and assessed the effectiveness of the iVac system and passive ultrasonic irrigation (PUI) compared to conventional irrigation for removing calcium hydroxide-based intracanal medication using micro-computed tomography (microCT). METHODS: Steel and titanium inserts from the Ultra X system (blue, silver, gold) and the polymer insert from the iVac system were analyzed using vibroacoustography in Ultra X (45 kHz) and Woodpecker (40 kHz) ultrasonic devices. Inserts were positioned in a water tank for standardized placement. The excitation signal was generated using MATLAB software, and power measurements were conducted in duplicate. After root canal preparation in 30 teeth, calcium hydroxide-based medication was applied and scanned using microCT. After 7 days, initial removal was performed with a K-file (number 45.02), followed by three protocols: conventional irrigation, PUI, or iVac. A second microCT scan was performed. Analysis of variance and Tukey's tests compared ultrasonic power and residual medication volume. RESULTS: The steel, titanium, and polymer inserts exhibited frequencies from 37 to 45 kHz. The iVac group (99.8 ± 0.3%) achieved greater medication removal than the PUI group (97.3 ± 2.9%), both significantly superior to conventional irrigation (73.0 ± 9.4%) (P < .05). CONCLUSION: The ultrasonic power of the inserts operated within the expected frequency range. The iVac system demonstrated superior effectiveness in removing intracanal medication.