刃部几何形态(非平面、平面或混合)与冶金成分对镍钛根管器械机械性能的影响:一项多方法实验室研究

The impact of blade geometry (nonflat, flat or hybrid) and metallurgical composition on the mechanical performance of NiTi endodontic instruments: A multimethod laboratory study.
PubMed 文献 · Int Endod J · 2025-05-03

本实验室研究纳入207支镍钛根管器械(常规刃n=69、平面刃n=69、混合刃n=69),评估几何设计、合金成分、相变温度及循环疲劳、扭转、弯曲、屈曲、切割效率和显微硬度。常规刃CC One Blue在断裂时间、最大扭矩和屈曲强度上表现更优;平面刃Platinum V.EU显微硬度更高但机械性能较差。

要点

  • 常规非平面刃器械在断裂时间、最大扭矩和屈曲强度上表现更优。
  • 平面刃Platinum V.EU显微硬度更高,但柔韧性、切割效率和机械性能较差。
  • 同一批次内Platinum V.EU和Flash存在设计不一致。

核心结论

  • 总体而言,具有平面侧(Platinum V.EU)或混合(Flash)活动刃的器械机械性能劣于常规非平面器械(CC One Blue)。

研究设计与样本

  • 实验室研究,选取207支长度25 mm的镍钛根管器械,按刃部设计分为常规刃(n=69,CC One Blue,尺寸25/0.08v)、平面刃(n=69,Platinum V.EU,尺寸25/0.06)和混合刃(n=69,Flash Endo Power,尺寸25/0.06v)三组。
  • 采用扫描电子显微镜评估几何设计、能量色散X射线光谱评估合金元素组成、差示扫描量热法评估相变温度,并测试循环疲劳抗力、扭转抗力、弯曲抗力、屈曲抗力、切割效率和显微硬度。
  • 统计显著性采用单因素方差分析和Kruskal-Wallis检验(α=5%)。

主要发现

  • Platinum V.EU和Flash器械在同一批次内存在设计不一致,包括非标准定位和平面侧长度变异。
  • 所有器械均由镍钛等原子比合金构成;20°C时Flash呈R相与奥氏体混合排列,36°C时完全转为奥氏体,而CC One Blue和Platinum V.EU在20°C呈完全R相、36°C保持R相与奥氏体混合排列。
  • CC One Blue在断裂时间(156±34 s)、最大扭矩(1.5 N·cm)和屈曲强度(372±31 gf)上表现更优(p<.0001),最大旋转角度无差异(p=.602)。
  • 柔韧性方面,Flash(328 gf)和CC One Blue(341 gf)优于Platinum V.EU(376 gf)(p=.006);切割效率方面,Flash(121 gf)和CC One Blue(137 gf)优于Platinum V.EU(253 gf)(p<.0001)。
  • Platinum V.EU显微硬度(386±45 HVN)显著高于CC One Blue和Flash(p=.0340)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIM: To evaluate the influence of blade design (conventional, flat and hybrid) and metallurgical properties on the mechanical performance of nickel-titanium endodontic instruments. METHODOLOGY: Two hundred and seven NiTi instruments (25 mm in length) with three different blade designs were selected for analysis: conventional (n = 69, CC One Blue, size 25/0.08v), flat (n = 69, Platinum V.EU, size 25/0.06) and hybrid (n = 69, Flash Endo Power, size 25/0.06v). The instruments were evaluated regarding geometric design (scanning electron microscopy), alloy elements composition (energy-dispersive X-ray spectroscopy) and phase transformation temperatures (differential scanning calorimetry). Additionally, their mechanical behaviour was investigated by testing cyclic fatigue resistance, torsional resistance, bending resistance, buckling resistance, cutting efficiency and microhardness. Statistical significance was determined using One-Way anova and Kruskal-Wallis tests (α = 5%). RESULTS: Platinum V.EU and Flash instruments exhibited design inconsistencies within the same lot, including nonstandard positioning and variations in the length of the flat side. All instruments were composed of a nickel-titanium alloy with equiatomic ratios of nickel and titanium. At 20°C, Flash instruments exhibited a mixed R-phase and austenitic arrangement, transitioning fully to austenitic at 36°C, while CC One Blue and Platinum V.EU displayed a complete R-phase at 20°C and retained a mixed R-phase and austenitic arrangement at 36°C. The CC One Blue exhibited superior performance in time to fracture (156 ± 34 s), maximum torque (1.5 N·cm) and buckling strength (372 ± 31 gf) (p < .0001), while no differences were found in maximum rotation angle (p = .602). In terms of flexibility, the Flash (328 gf) and CC One Blue (341 gf) outperformed the Platinum V.EU (376 gf) (p = .006). Flash (121 gf) and CC One Blue (137 gf) also outperformed Platinum V.EU (253 gf) in terms of cutting efficiency (p < .0001). Conversely, the Platinum V.EU demonstrated significantly higher microhardness (386 ± 45 HVN) compared to CC One Blue and Flash (p = .0340). CONCLUSIONS: Overall, instruments featuring either flat-side (Platinum V.EU) or hybrid (Flash) active blades demonstrated inferior mechanical performance compared to the conventional nonflat instrument (CC One Blue).

来源:Int Endod J,查看原文。版权归原作者所有。

引用

  1. Silva EJNL, Martins JNR, Cardoso TO, et al. The impact of blade geometry (nonflat, flat or hybrid) and metallurgical composition on the mechanical performance of NiTi endodontic instruments: A multimethod laboratory study. Int Endod J. 2025. PMID: 40317569. doi:10.1111/iej.14249