文献速递口腔修复与种植

人类颌骨牙种植体初期稳定性:实验与有限元分析

Primary Stability of Dental Implants in Human Jawbones: Experiments & FE Analyses.
PubMed 文献 · Clin Oral Implants Res · 2024-12-13

研究采用生物力学测试与均匀化有限元分析,对来自两具供体颌骨的35份骨样本(最终评估22份)测量种植体初期稳定性,发现模拟极限力与实验极限力的相关性(R²=0.85)高于骨体积分数、植入扭矩和共振频率分析,所有变量与实验极限力均呈显著线性相关(p<0.01)。

要点

  • 模拟极限力与实验极限力相关性R²=0.85,高于骨体积分数(R²=0.61)、植入扭矩(R²=0.50)和共振频率分析(R²=0.38)。
  • 所有变量与实验极限力均呈统计学显著线性相关(p<0.01)。
  • 均匀化有限元分析需要预先的μCT重建,目前受数值收敛问题限制。

核心结论

  • 模拟极限力是比植入扭矩和共振频率分析更可靠、更客观的初期稳定性指标。
  • 术前均匀化有限元分析在估计初期稳定性方面比现有技术更准确,是有前景的工具,但需要预先的μCT重建,且目前受数值收敛问题限制。

研究设计与样本

  • 研究对来自两具供体颌骨的35份骨样本进行检测,最终22份纳入初期稳定性评估。
  • 样本经μCT扫描后由两名经验丰富的外科医生植入种植体,评估基于μCT的骨体积分数、植入扭矩和共振频率分析。
  • 进行力学测试测量实验极限力作为初期稳定性的客观指标,同时用均匀化有限元模拟估计模拟极限力。

主要发现

  • 模拟极限力与实验极限力的相关性R²=0.85,高于骨体积分数与实验极限力(R²=0.61)、植入扭矩与实验极限力(R²=0.50)以及共振频率分析与实验极限力(R²=0.38)。
  • 所有变量与实验极限力均呈统计学显著线性相关(p<0.01)。

临床怎么用

  • 原文结论认为模拟极限力是比植入扭矩和共振频率分析更可靠、更客观的初期稳定性指标。
  • 原文指出术前均匀化有限元分析在估计初期稳定性方面比现有技术更准确,是有前景的工具。

局限性

  • 均匀化有限元分析需要预先的μCT重建。
  • 均匀化有限元分析目前受数值收敛问题限制。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Primary stability (PS) is a key factor for promoting osseointegration and long-term success of dental implants particularly for immediate loading protocols. Beyond the current assessments of PS, an accurate pre-operative evaluation of PS would contribute to the improvement of surgical planning and treatment outcome. This study used biomechanical testing and homogenized finite element (hFE) analysis to objectively measure PS in the laboratory, and digitally estimate PS from prior μCT reconstructions. MATERIAL AND METHODS: Thirty-five bone samples extracted from the jaws of two donors were examined. Twenty-two were finally evaluated for PS. After scanning of the samples with μCT, implants were inserted by two experienced surgeons, and various metrics such as μCT-based bone volume fraction (BV/TV), insertion torque (IT), and resonance frequency analysis (RFA) were assessed to determine PS. Mechanical tests were conducted to measure ultimate force (UFexp) as an objective indicator of PS while the hFE simulations were performed to estimate this same ultimate force (UFsim). RESULTS: Higher correlation was found between UFsim and UFexp (R2 = 0.85) than between BV/TV and UFexp (R2 = 0.61), IT and UFexp (R2 = 0.50), and RFA and UFexp (R2 = 0.38). All variables demonstrated a statistically significant linear correlation with UFexp (p < 0.01). CONCLUSION: UFsim turns out to be a more reliable and objective indicator of PS than IT and RFA. The hFE analysis requires prior μCT reconstructions and is currently limited by numerical convergence problems. Despite these limitations, pre-operative hFE analysis emerges as a promising tool with a higher accuracy for estimation of PS than state of care techniques.

来源:Clin Oral Implants Res,查看原文。版权归原作者所有。

引用

  1. Wili P, Rauber C, Saade A, et al. Primary Stability of Dental Implants in Human Jawbones: Experiments & FE Analyses. Clin Oral Implants Res. 2024. PMID: 39670698. doi:10.1111/clr.14386

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。