氧化锆种植体数量与CBCT重建厚度对种植体周围骨缺损检测的影响:一项离体诊断准确性研究

Do the number of zirconia implants and the thickness of CBCT image reconstruction affect the detection of peri-implant bone defect? A diagnostic accuracy ex vivo study.
PubMed 文献 · Clin Oral Implants Res · 2024-07-27

结果显示各因素对诊断无显著影响(p > .05),单颗种植体在2 mm重建厚度时诊断性能较高(AUC = 0.88,敏感度 = 0.68,特异度 = 0.94)。结论认为相邻氧化锆种植体的存在及重建厚度变化不影响CBCT对3 mm颊侧种植体周围骨缺损的检测。

要点

  • 相邻氧化锆种植体存在与否不影响CBCT对3 mm颊侧骨缺损的检测。
  • 0.125 mm、1 mm、2 mm重建厚度间诊断性能无显著差异(p > .05)。
  • 单颗种植体配合2 mm重建厚度时AUC达0.88,敏感度0.68,特异度0.94。

核心结论

  • 相邻氧化锆种植体的存在及重建厚度变化不影响CBCT对3 mm颊侧种植体周围骨缺损的检测。

研究设计与样本

  • CBCT扫描参数为90 kVp、6.3 mA、5×5 cm FOV、0.125 mm³体素。
  • 比较单颗与两颗氧化锆种植体及0.125 mm、1 mm、2 mm三种重建厚度。
  • 由5名口腔颌面放射科医师采用5分法评估,使用双因素方差分析(α = .05)。

主要发现

  • 所研究的因素对种植体周围骨缺损诊断无显著影响(p > .05)。
  • 单颗种植体时诊断性能较高,尤其2 mm重建厚度下AUC = 0.88、敏感度 = 0.68、特异度 = 0.94。
  • 两颗种植体时结果较温和但差异无统计学意义(AUC = 0.80、敏感度 = 0.58、特异度 = 0.82)。

临床怎么用

  • CBCT评估3 mm颊侧种植体周围骨缺损时,相邻氧化锆种植体不构成显著干扰。
  • 重建厚度在0.125 mm至2 mm范围内对诊断准确性无显著影响。

局限性

  • 离体研究使用干燥人下颌骨,结果外推至临床活体需谨慎。
  • 仅评估3 mm颊侧骨缺损,其他尺寸和位置未涉及。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: To evaluate the influence of multiplanar reconstruction thickness on the detection of peri-implant bone defects with a standalone zirconia implant and compare it to when another implant is in the vicinity using cone-beam computed tomography (CBCT). MATERIALS AND METHODS: Five dry human mandibles were used to create twenty implant sites in the second premolar and first molar regions. The OP300 Maxio was used to acquire CBCT images (90 kVp, 6.3 mA, 5 × 5 cm FOV, and 0.125 mm3 voxel size) before and after creating 3 mm peri-implant bone defects in the buccal aspect of the premolar region. Half of the scans featured a single zirconia implant in the premolar region, while the others had two implants in the premolar and molar regions. Three reconstruction thicknesses (0.125 mm, 1 mm, and 2 mm) were considered for the multiplanar reconstruction analyses. Five oral and maxillofacial radiologists assessed the detection of peri-implant bone defects using a 5-point scale. Diagnostic parameters were calculated and compared using Two-way ANOVA (α = .05). RESULTS: The studied factors showed no significant influence on the diagnosis of peri-implant bone defects (p > .05). Diagnostic performance was notably higher with a single implant, especially with a 2-mm reconstruction thickness (AUC = 0.88, sensitivity = 0.68, specificity = 0.94). Although the differences were not statistically significant, the results were more modest when two implants were present (AUC = 0.80, sensitivity = 0.58, specificity = 0.82). CONCLUSIONS: The presence of an adjacent zirconia implant and variations in reconstruction thickness did not influence the detection of 3 mm buccal peri-implant bone defects on CBCT images.

来源:Clin Oral Implants Res,查看原文。版权归原作者所有。