不同口内扫描仪对无牙颌上下颌咬合记录精度的影响

Effect of Different Intraoral Scanners on the Accuracy of Bite Registration in Edentulous Maxillary and Mandibular Arches.
PubMed 文献 · J Dent · 2024-05-10

本研究采用体外模型评估四种口内扫描仪(IOS)在无牙颌上颌和下颌弓的咬合记录精度。在上颌和下颌无牙颌模型上对称植入四颗种植体后,使用Primescan、Trios 3、Trios 4和Medit i500进行数字印模和咬合记录,每种IOS样本量为10。结果显示不同IOS间的正确度和精密度存在显著差异(p<.05),仅Primescan达到临床可接受的精度。

要点

  • Primescan在无牙颌咬合记录中精度最高,优于Medit i500、Trios 3和Trios 4。
  • 仅Primescan达到临床可接受的颌间记录精度。
  • 不同IOS类型影响无牙颌咬合记录的精度。

核心结论

  • 在本研究局限性内,IOS类型影响体外设计中颌间咬合记录的精度。
  • 仅Primescan对无牙颌弓的颌间记录达到临床可接受的精度。

研究设计与样本

  • 体外研究,基于Frasaco模型形状用Asiga Max UV 3D打印机打印无牙颌上颌和下颌模型。
  • 每个模型对称植入四颗Straumann BLT RC种植体。
  • 使用Primescan、Trios 3、Trios 4和Medit i500四种口内扫描仪进行数字印模,每种IOS样本量为10。

主要发现

  • 不同IOS间的正确度和精密度存在显著差异(p<.05),但Medit i500与Trios 3之间无显著差异(p>.05)。
  • 精度排序为Primescan最准确,其次为Medit i500、Trios 3、Trios 4。

关键数值与推荐

  • 不同IOS间的正确度和精密度比较,除Medit i500与Trios 3外,均显示显著差异(p<.05)。
  • 仅Primescan达到临床可接受的颌间记录精度。

临床怎么用

  • 选择口内扫描仪进行无牙颌咬合记录时,应考虑其精度差异,Primescan可能提供更可靠的颌间记录。

局限性

  • 本研究为体外设计,结论受此局限性限制。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The objective of this study was to use in vitro models to examine the bite registration accuracy of four different intraoral scanners (IOS) for edentulous maxillary and mandibular arches. The objective was to assess the trueness and precision of the IOS and determine if there were significant differences between them. METHODS: An Asiga Max UV 3D printer was used to print maxillary and mandibular edentulous models based on the shape of Frasaco models (artificial dental arch models). Four dental implants were placed symmetrically in both models using Straumann BLT RC implants. Digital impressions were taken with Primescan, Trios 3, Trios 4, and Medit i500 intraoral scanners (n = 10 for each IOS). Digital bite registrations were made, and scanning data was exported in STL format. The accuracy of the interarch distance (the distance between the metrological spheres attached to the mandibular and maxillary models) was estimated for each IOS. RESULTS: The results showed significant differences in trueness and precision between different IOS (p <.05), except Medit i500 and Trios 3 (p >.05). Primescan provided the most accurate results, followed by Medit i500, Trios 3, and Trios 4, respectively. CONCLUSIONS: within the limitations of this study, the IOS type affects the accuracy of interocclusal bite registration in in vitro design. Only Primescan achieved clinically acceptable accuracy for the interocclusal recording of edentulous arches. CLINICAL RELEVANCE: The comparison of the accuracy of bite registration between different intraoral scanners will help increase the efficiency of the clinical application of digitalized interarch registration.