文献速递口腔修复与种植

新型锥形组织水平种植体与平行壁种植体初期稳定性及骨结合对比

Primary stability and osseointegration comparing a novel tapered design tissue-level implant with a parallel design tissue-level implant. An experimental in vivo study.
PubMed 文献 · Clin Oral Implants Res · 2024-05-31

本研究在小型猪下颌骨植入新型锥形双螺纹自攻组织水平种植体(TLC)与成熟平行壁组织水平种植体(TL),比较3、6、12周初期及继发稳定性。TLC最大植入扭矩显著更高(57.83±24.73 Ncm vs 22.62±23.16 Ncm,p<.001),ISQ无差异;各时间点BIC相当,12周时TLC的fBIC显著更靠冠方(0.31±0.83 mm vs -0.22±0.85 mm,p=.027)。结论:TLC初期稳定性更优,12周时牙槽嵴顶骨高度维持更好。

要点

  • TLC种植体最大植入扭矩显著高于TL(p<.001),初期稳定性更优。
  • 两种种植体ISQ及各时间点BIC无显著差异。
  • 12周时TLC的fBIC显著更靠冠方,提示嵴顶骨高度维持更好。

核心结论

  • 新型锥形组织水平种植体初期稳定性更优,12周时牙槽嵴顶骨高度维持整体更好。

研究设计与样本

  • 实验性体内研究,将TLC(每时间点n=10)与TL(每时间点n=10)种植体植入小型猪下颌骨,埋入式愈合3、6、12周。

主要发现

  • TLC最大植入扭矩为57.83±24.73 Ncm,显著高于TL的22.62±23.16 Ncm(p<.001)。
  • TL与TLC平均ISQ分别为75.00±6.70与75.40±3.20,差异无统计学意义(p=.988)。
  • 各评估时间点两种种植体BIC相当。
  • 12周时TLC的fBIC为0.31±0.83 mm,显著较TL的-0.22±0.85 mm更靠冠方(p=.027)。

临床怎么用

  • 在需较高初期稳定性的病例中,可考虑选用该锥形自攻组织水平设计种植体。
  • 该设计可能有助于12周时维持牙槽嵴顶骨高度,但需临床研究验证。

局限性

  • 研究为小型猪动物实验,样本量有限,结论外推至临床需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The aim of the present study was to compare a novel tapered, double-threaded self-tapping tissue-Level design implant (TLC) to a well-established parallel walled tissue-level (TL) implant in terms of primary and secondary stability over time. MATERIALS AND METHODS: Test TLC (n = 10/per timepoint) and control TL (n = 10/per timepoint) implants were placed in the mandible of minipigs and left for submerged healing for 3, 6, and 12 weeks. Maximum insertion torque and implant stability quotient (ISQ) were measured for each implant at placement. Osseointegration and cortical bone maintenance were histologically evaluated by measuring total bone-to-implant contact (BIC) and first bone-to-implant contact (fBIC). RESULTS: A significantly higher maximum insertion torque was measured for the test implant TLC compared to the control TL implant (57.83 ± 24.73 Ncm and 22.62 ± 23.16 Ncm, respectively; p < .001). The mean ISQ values were comparable between the two implant types (75.00 ± 6.70 for TL compared to 75.40 ± 3.20 for TLC, p = .988). BIC was comparable between both implant types at each of the evaluated time points. The fBIC was found to be significantly more coronal at 12 weeks for the TLC implant compared to the TL implant (0.31 ± 0.83 mm for TLC compared to -0.22 ± 0.85 for TL, p = .027). CONCLUSION: The novel tapered tissue level design implant showed improved primary stability and an overall improved crestal bone height maintenance compared to the parallel walled design at 12 weeks.

来源:Clin Oral Implants Res,查看原文。版权归原作者所有。

引用

  1. Gill T, Kühl S, Rawlinson S, et al. Primary stability and osseointegration comparing a novel tapered design tissue-level implant with a parallel design tissue-level implant. An experimental in vivo study. Clin Oral Implants Res. 2024. PMID: 38819108. doi:10.1111/clr.14301

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。