温度敏感镍钛器械形态变化与根管内实时温度动态研究
研究在22至45°C水浴中实时观察XP-Endo Shaper、Finisher和Finisher-R的形态与长度变化,并在下颌磨牙非手术根管治疗中测量根管内温度。三种器械形变起始与最佳转化温度不同,完全转化后长度分别平均缩短0.43、1.07和1.15 mm;无论冲洗液22或45°C,3分钟后根管内温度仅达32.9±0.8和33.2±1.0°C,提示体外测试需考虑冲洗液温度。
要点
- XP-Endo Shaper形变始于31.5±2.0°C,最佳转化于35.1±1.0°C。
- Finisher与Finisher-R形变始于29.2±1.9和26.9±2.2°C,最佳转化于33.9±1.4和32.7±1.7°C。
- 完全转化后三种器械长度平均缩短0.43±0.23、1.07±0.22、1.15±0.22 mm。
核心结论
- 三种温度敏感镍钛器械在不同温度下呈现不同的形态与长度变化。
- 无论冲洗液温度为22或45°C,根管内温度达到平衡后均低于冲洗液温度。
- 体外测试牙髓器械时应考虑冲洗液温度的影响。
研究设计与样本
- 体外实验:XP-Endo Shaper、Finisher和Finisher-R在22至45°C控温水浴中加热,实时拍摄并每10秒提取帧测量长度与形态变化。
- 体内实验:对接受非手术根管治疗患者的下颌磨牙远中根,用K型热电偶探头于根中1/3测量根管内温度。
- 冲洗液分别设为室温22°C或加热至45°C,录像记录口内与根管内温度180秒,每5秒间隔绘图。
主要发现
- XP-Endo Shaper形变始于31.5±2.0°C,最佳转化于35.1±1.0°C。
- Finisher形变始于29.2±1.9°C,最佳转化于33.9±1.4°C;Finisher-R形变始于26.9±2.2°C,最佳转化于32.7±1.7°C。
- 完全转化后XP-Endo Shaper、Finisher和Finisher-R长度平均分别缩短0.43±0.23、1.07±0.22和1.15±0.22 mm。
- 冲洗液22或45°C作用3分钟后,根管内温度分别达32.9±0.8和33.2±1.0°C。
关键数值与推荐
- XP-Endo Shaper形变起始温度31.5±2.0°C,最佳转化温度35.1±1.0°C。
- Finisher形变起始温度29.2±1.9°C,最佳转化温度33.9±1.4°C。
- Finisher-R形变起始温度26.9±2.2°C,最佳转化温度32.7±1.7°C。
- 完全转化后长度缩短:XP-Endo Shaper 0.43±0.23 mm、Finisher 1.07±0.22 mm、Finisher-R 1.15±0.22 mm。
- 冲洗液22或45°C作用3分钟后根管内温度分别为32.9±0.8和33.2±1.0°C。
临床怎么用
- 使用温度敏感镍钛器械时需知晓其形变起始与最佳转化温度,以评估临床操作中的形态适应行为。
- 体外评价牙髓器械时应模拟根管内实际温度或考虑冲洗液温度,避免仅凭水温推断器械性能。
- 冲洗液加热至45°C后3分钟根管内温度仍约33°C,提示临床等待温度平衡的时间与效果需纳入考量。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIM: This study was twofold: (i) it aimed to investigate the morphometric changes of three temperature-sensitive nickel-titanium (NiTi) instruments at different temperatures, and (ii) to conduct an in vivo real-time analysis of intracanal temperature changes. METHODS: Changes in the shape and length of XP-Endo Shaper, XP-Endo Finisher, and XP-Endo Finisher-R were evaluated in real time whilst heated in a temperature-controlled water bath from 22 to 45°C. Instruments were fixed to a laminated water-resistant 1 mm graph paper attached to a stone block. Instruments were imaged whilst subjected to increasing temperature using a digital camera attached to an operating microscope. From recorded videos, still frames were extracted at 10-s intervals and changes in the length and shape of each instrument were measured and changes were plotted against time. Moreover, the intracanal temperature of distal roots of lower molars was measured in vivo for patients attending the clinic for non-surgical root canal treatments. The temperature was measured using a K-type thermocouple probe inserted into the mid-root level after irrigating the canal with a solution set at room temperature (22°C) or heated to 45°C. The intraoral and intracanal temperatures were recorded using a video camera for 180 s at 5-s intervals to plot the change in the intraoral and intracanal temperature, after both irrigation solution temperatures, with time. RESULTS: The shape transformation of XP-Endo Shaper began at 31.5 ± 2.0°C and reached its optimal transformation at 35.1 ± 1.0°C. For the Finisher and Finisher-R, shape transformations began at 29.2 ± 1.9 and 26.9 ± 2.2°C reaching the optimal transformation at 33.9 ± 1.4 and 32.7 ± 1.7°C, respectively. The average decreases in lengths of XP-Endo Shaper, Finisher, and Finisher-R after full transformation were 0.43 ± 0.23, 1.07 ± 0.22, and 1.15 ± 0.22 mm, respectively. The intracanal temperature reached 32.9 ± 0.8 and 33.2 ± 1.0°C after 3 min of application of irrigation solutions set at 22 or 45°C, respectively. CONCLUSION: The tested instruments exhibited diverse changes in their shapes and lengths at varying temperatures. Despite the temperature of the irrigation solution, the intracanal temperature consistently remained lower than the intracanal temperature once equilibrium was reached. This highlights the importance of considering the temperature of irrigation solution during in vitro testing of endodontic instruments.
引用
- Atmeh AR, Al Omari T, AlQifari AF, et al. Exploring morphometric transformations in temperature-sensitive NiTi instruments alongside real-time intracanal temperature dynamics. Int Endod J. 2024. PMID: 38804711. doi:10.1111/iej.14100
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。