模板引导种植体植入精度:不同制作方法的体外比较

Accuracy of template-based guided dental implant placement-An in vitro comparison of different manufacturing methods.
PubMed 文献 · Clin Oral Implants Res · 2024-04-17

本研究在体外模型上比较增材3D打印、减材研磨和传统技工室制作三种手术导板对单牙种植体位置三维传递精度的影响。每种方法制作10个导板,通过CBCT扫描评估种植体肩部和根尖偏差及角度偏差。结果显示不同制作方法间无统计学差异,所有方法的种植体偏差均未超出公认安全范围(1.5-2.0 mm)。

要点

  • 三种导板制作方法(3D打印、研磨、传统)的种植体植入精度无统计学差异。
  • 种植体肩部垂直偏差最大达1.04 mm(3D打印组),水平偏差较小。
  • 所有方法的角度偏差中位数在2.1°至3.3°之间,均未超出安全偏差范围。

核心结论

  • 在体外环境下,手术导板制作方法不影响种植体植入精度。
  • 所有制作方法获得的种植体位置均未超出公认的安全偏差范围(1.5-2.0 mm)。

研究设计与样本

  • 体外研究,针对单牙缺隙(#26)数字化规划种植体位置。
  • 每个导板用于模型复制体上的模板引导种植体植入。
  • 通过CBCT扫描重叠植入模型并确定种植体位置,测量肩部和根尖偏差及角度。

主要发现

  • 种植体肩部垂直方向偏差主导位置偏差,3D打印组最大达1.04 mm(IQR 0.28 mm)。
  • 水平偏差较小,近远中向最大达0.38 mm(IQR 0.36 mm,传统技工室组)。
  • 种植体根尖水平偏差较肩部明显增加,尤其在近远中方向。
  • 角度偏差中位数:3D打印组2.1°(IQR 2.0 mm,最大4.2°),研磨组3.3°(IQR 1.9 mm,最大5.3°)。
  • 不同制作方法间无统计学差异(Kruskal-Wallis检验,p = .05)。

关键数值与推荐

  • 公认的安全偏差范围为1.5-2.0 mm(原文结论中提及)。

临床怎么用

  • 在体外条件下,三种导板制作方法均可获得可接受的种植体植入精度。
  • 临床选择导板制作方法时可综合考虑成本、可及性等因素,精度差异不显著。

局限性

  • 研究为体外环境,结果不能直接外推至临床口内实际情况。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: This study investigates effects of surgical guide manufacturing process on 3D transfer accuracy of planned dental implant position, using three production methods: additive 3D-printed (3DF), subtractive milled (MF), and analog laboratory fabricated templates (LF). MATERIAL AND METHODS: Implant position for a single-tooth gap (#26) planned digitally. 3DF and MF templates were designed digitally, while LF templates were analogously created. For each manufacturing type, 10 surgical guides were fabricated. Each guide was used for template-guided implant placement in model replicas. For evaluation of implant placement, cone beam computed tomography scans of all implanted models were superimposed, and implant positions were determined. Deviations at implant shoulder/apex were measured, and median and inter-quartile range (IQR) were determined for mesio-distal, oro-facial, coronal apico, 3D spaces, and angles. RESULTS: At implant shoulder, vertical components dominated position deviations (up to 1.04 mm, IQR 0.28 mm for 3DF). Horizontal deviations were much lower (mesio-distally up to 0.38 mm, IQR 0.36 mm (LF)). Implant apex shows similar vertical deviations, while horizontal deviations clearly increased compared to shoulder, especially in mesio-distal direction. Median angular deviations were between 2.1° (IQR 2.0 mm, max. 4.2°) for 3DF and 3.3° (IQR 1.9 mm, max. 5.3°) for MF. No statistical differences were found between manufacturing types (Kruskal-Wallis test, p = .05). CONCLUSIONS: The study showed the method of implant guide fabrication did not affect the accuracy of implant placement within the limits of an in vitro environment. All methods resulted in implant placement which did not exceed the accepted safety deviation envelope (1.5-2.0 mm).

来源:Clin Oral Implants Res,查看原文。版权归原作者所有。