自主机器人辅助全牙弓种植手术的1年前瞻性临床评价

Clinical evaluation of autonomous robotic-assisted full-arch implant surgery: A 1-year prospective clinical study.
PubMed 文献 · Clin Oral Implants Res · 2024-02-06

结果显示全局平台偏差、根尖偏差和角度偏差分别为0.53(0.19)mm、0.58(0.17)mm和1.83(0.82)°,随访1年牙周参数、颊侧骨厚度和面部垂直骨壁峰值保持稳定。结论认为ADIR系统位置精度优异,全牙弓种植术后1年临床性能良好。

要点

  • ADIR系统全牙弓种植全局平台偏差0.53mm、根尖偏差0.58mm、角度偏差1.83°。
  • 颌骨、种植位点、患者顺序、种植系统和种植体长度对偏差无显著影响。
  • 术后1年牙周参数、颊侧骨厚度和面部垂直骨壁峰值保持稳定。

核心结论

  • ADIR系统在全牙弓种植手术中表现出优异的位置精度。
  • 全牙弓种植术后1年随访显示ADIR系统可获得良好的临床性能。
  • 仍需更多临床证据为ADIR系统在全牙弓种植手术中的应用提供指南。

研究设计与样本

  • 使用自主牙种植机器人(ADIR)系统进行全牙弓种植手术。
  • 术后计算全局平台偏差、全局根尖偏差和全局角度偏差,并分析颌骨、种植位点、患者顺序、种植系统和种植体长度的影响。
  • 记录手术时长,术后3个月和1年随访,记录牙周参数、颊侧骨厚度(BBT)和面部垂直骨壁峰值(IP-FC)。

主要发现

  • 102枚种植体的平均(SD)全局平台偏差为0.53(0.19)mm,全局根尖偏差为0.58(0.17)mm,全局角度偏差为1.83(0.82)°。
  • 下颌与上颌之间的偏差差异无统计学意义(p > .05)。
  • 颌骨、种植位点、患者顺序、种植系统和种植体长度对偏差均无统计学显著差异(p > .05)。
  • 牙周参数、颊侧骨厚度(BBT)和面部垂直骨壁峰值(IP-FC)在1年随访期间保持稳定。

临床怎么用

  • ADIR系统可用于全牙弓种植手术,具有较高的种植体位置精度。
  • 术后1年随访显示牙周参数和骨壁指标稳定,支持其短期临床效果。
  • 仍需更多临床证据以制定ADIR系统在全牙弓种植手术中的应用指南。

局限性

  • 随访时间仅为1年,长期效果尚不明确。
  • 作者指出需要额外临床证据来制定ADIR系统在全牙弓种植手术中的实施指南。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: This prospective clinical study aimed to evaluate the accuracy and 1-year clinical follow-up performance of dental implant placement with an autonomous dental implant robot (ADIR) system in full-arch implant surgery. MATERIALS AND METHODS: Twelve patients with edentulous arches or final dentition received 102 implants using the ADIR system. Global platform deviation, global apex deviation, and global angular deviation between the planned and actual implants were calculated after surgery. Data were statistically analyzed for factors including jaws, implant positions, patient sequences, implant systems, and implant length. Surgery duration was recorded. Patients were followed for 3 months and 1 year after surgery. Periodontal parameters, buccal bone thickness (BBT), and facial vertical bone wall peak (IP-FC) were recorded. RESULTS: Among the 102 implants, the mean (SD) global platform deviation, global apex deviation, and global angular deviation were 0.53 (0.19) mm, 0.58 (0.17) mm, and 1.83 (0.82)°, respectively. The deviation differences between the mandible and maxilla did not show statistical significance (p > .05). No statistically significant differences were found for the jaws, implant positions, patient sequences, implant systems, and implant length to the deviations (p > .05). The periodontal parameters, the BBT, and IP-FC remained stable during 1-year follow-up. CONCLUSION: The ADIR system showed excellent positional accuracy. The 1-year follow-up after full-arch implant surgery indicated that the ADIR system could achieve promising clinical performance. Additional clinical evidence is requisite to furnish guidelines for the implementation of the ADIR system in full-arch implant surgery.