文献速递口腔修复与种植

3D打印p-cell结构氧化锆种植体的生物学性能

Biological Properties of 3D-Printed Zirconia Implants with p-Cell Structures.
PubMed 文献 · J Dent Res · 2024-02-19

本研究评估3D打印p-cell结构氧化锆种植体的生物学反应。实体样本密度、三点弯曲强度和加速老化性能与传统方法相当;p-cell结构促进细胞增殖、黏附和成骨相关蛋白表达;动物实验中p-cell组骨小梁更厚、更多、更致密。

要点

  • 3D打印实体氧化锆的密度、三点弯曲强度和加速老化性能与传统方法相当
  • p-cell结构促进细胞增殖、黏附和成骨相关蛋白表达
  • p-cell与对照种植体均能承受35 Ncm扭矩而不折断

核心结论

  • 3D打印p-cell结构氧化锆种植体展现出良好的生物学性能,为多孔氧化锆仿生种植体的进一步探索提供依据。

研究设计与样本

  • 研究评估3D打印制备的新型p-cell结构氧化锆种植体的生物学反应,包含材料性能测试、细胞实验、机械测试和动物实验。

主要发现

  • 实体氧化锆样本的密度、三点弯曲强度和加速老化性能与传统方法制备的标本相当。
  • 细胞实验显示p-cell结构促进细胞增殖、黏附和成骨相关蛋白表达。
  • p-cell和对照种植体均能承受35 Ncm扭矩而不折断。
  • p-cell和对照种植体的平均最大断裂载荷分别为1,222.429 ± 115.591 N和1,903.857 ± 250.673 N,远高于人类生理咀嚼力和平均最大咬合力。
  • 动物实验显示p-cell组种植体周围骨小梁较对照组显著更厚、更多且更致密。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Research on 3-dimensional (3D) printed porous zirconia-based dental implants is still in its infancy. This study aimed to evaluate the biological responses of novel zirconia implants with p-cell structures fabricated by 3D printing. The solid zirconia samples exhibited comparable density, 3-point flexural strength, and accelerated aging properties compared to specimens prepared previously by conventional methods. Cell-based experiments showed that the p-cell structure promoted cell proliferation, adhesion, and osteogenesis-related protein expression. Mechanical tests showed that both p-cell and control implants could withstand a torque of 35 Ncm without breaking. The mean maximum breaking loads of p-cell and control implants were 1,222.429 ± 115.591 N and 1,903.857 ± 250.673 N, respectively, which were much higher than the human physiological chewing force and human mean maximum occlusal force. An animal experiment showed that the bone trabeculae around the implants were significantly thicker, more numerous, and denser in the p-cell group than in the control group. This work could provide promising guidance for further exploring 3D printing techniques for porous zirconia bionic implants in dentistry.

来源:J Dent Res,查看原文。版权归原作者所有。

引用

  1. Qin W, Shen MH, Gan N, et al. Biological Properties of 3D-Printed Zirconia Implants with p-Cell Structures. J Dent Res. 2024. PMID: 38374666. doi:10.1177/00220345231222819

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。