增强现实头戴设备与动态导航系统用于上颌磨牙桩取出

Augmented Reality Head-Mounted Device and Dynamic Navigation System for Postremoval in Maxillary Molars.
PubMed 文献 · J Endod · 2024-02-17

本研究评估增强现实头戴设备(AR HMD)显示动态导航系统(DNS)用于上颌磨牙纤维桩取出的可行性,并与DNS技术比较。50颗上颌第一磨牙分为AR HMD+DNS组(n=25)和DNS组(n=25)。结果显示AR HMD+DNS比DNS更准确,全局冠根偏差和角度偏转显著更低(P < .05),均无意外发生,操作时间无显著差异(P > .05)。

要点

  • AR HMD显示DNS可安全用于上颌磨牙纤维桩取出。
  • AR HMD+DNS的全局冠根偏差和角度偏转显著低于DNS。
  • 两组操作时间无显著差异,均高效。

核心结论

  • 在体外研究局限内,AR HMD可安全显示DNS用于上颌磨牙纤维桩取出。
  • AR HMD提高了DNS的准确性。
  • AR HMD+DNS和DNS取出纤维桩均时间高效。

研究设计与样本

  • 体外研究,50颗上颌第一磨牙分为AR HMD+DNS组(n=25)和DNS组(n=25)。
  • 腭根用RelyX纤维桩和RelyX Unicem粘接,Paracore核build-up。
  • 术前术后拍摄锥形束CT(CBCT),用X-guide软件规划钻孔轨迹和深度。
  • AR HMD+DNS组用Microsoft HoloLens 2在术区显示DNS。
  • 计算三维偏差(全局冠根偏差GCD和全局根尖偏差GAD)和角度偏转AD,记录意外数量和操作时间。

主要发现

  • 所有样本(50/50)纤维桩均取出。
  • AR HMD+DNS比DNS更准确,GCD、GAD偏差和AD显著更低(P < .05)。
  • 无意外发生。
  • AR HMD+DNS与DNS操作时间无显著差异(P > .05)。

临床怎么用

  • AR HMD可安全显示DNS用于上颌磨牙纤维桩取出。
  • AR HMD+DNS可提高DNS准确性。
  • AR HMD+DNS和DNS均时间高效。

局限性

  • 本研究为体外研究。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: This study evaluates the feasibility of an augmented reality (AR) head-mounted device (HMD) displaying a dynamic navigation system (DNS) in the surgical site for fiber postremoval in maxillary molars and compares it to the DNS technique. METHODS: Fifty maxillary first molars were divided into 2 groups: AR HMD + DNS (n = 25) and DNS (n = 25). The palatal canal was restored with RelyX fiber post (3M ESPE) luted with RelyX Unicem (3M ESPE). A core buildup was performed using Paracore (Coltene/Whaledent). Cone beam computed tomography (CBCT) scans were taken before and after postremoval. The drilling trajectory and depth were planned under X-guide software (X-Nav Technologies, Lansdale, PA). For the AR HMD + DNS group, the AR HMD (Microsoft HoloLens 2) displayed the DNS in the surgical site. The three dimensional (3D) deviations (Global coronal deviation [GCD] and global apical deviation [GAD]) and angular deflection (AD) were calculated. The number of mishaps and operating time were recorded. RESULTS: Fiber post was removed from all samples (50/50). The AR HMD + DNS was more accurate than DNS, showing significantly lower GCD and GAD deviations and AD (P < .05). No mishap was detected. The AR HMD + DNS was as efficient in time as DNS (P > .05). CONCLUSIONS: Within the limitations of this in vitro study, the AR HMD can safely display DNS in the surgical site for fiber post-removal in maxillary molars. AR HMD improved the DNS accuracy. Both AR HMD + DNS and DNS were time-efficient for fiber postremoval in maxillary molars.

来源:J Endod,查看原文。版权归原作者所有。