文献速递牙体牙髓病

激光激活荡洗与机械激活技术对根管峡部清洁效率的比较分析

Comparative Analysis of Irrigation Techniques for Cleaning Efficiency in Isthmus Structures.
PubMed 文献 · J Endod · 2024-02-20

本研究在3D打印双弯根管模型中比较激光激活荡洗(LAI)与多种机械激活技术去除峡部模拟生物膜水凝胶的效果。结果显示LAI对整个根管系统的清洁效果最佳,其次为EDDY(SAI-E);在峡部区域,SAI-E和传统针头冲洗(NI)与LAI效果相当。

要点

  • 激光激活荡洗(LAI)对整个根管系统的清洁效率优于其他技术。
  • 在根管峡部,EDDY(SAI-E)和传统针头冲洗(NI)与LAI效果相当。
  • 超声激活荡洗(UAI)在峡部清洁中表现最差。

核心结论

  • 激光激活荡洗(LAI)对整个根管系统的清洁效果优于其他冲洗技术。
  • 在峡部区域,EDDY(SAI-E)和传统针头冲洗(NI)与LAI效果相当。

研究设计与样本

  • 研究采用3D打印双弯根管模型,共纳入n=20个样本进行冲洗方案比较。

主要发现

  • LAI对整个根管系统的水凝胶去除率最高(P < .05),其次为SAI-E。
  • 冠部峡部:LAI、SAI-E、NI和MA之间无显著差异(P > .05),SAI-EA和UAI效果较差(P < .05)。
  • 根尖峡部:所有技术均优于UAI(P < .05),其中LAI、SAI-E和NI效果最佳(P < .05)。

关键数值与推荐

  • 冲洗方案均为连续冲洗下蒸馏水,每个根管3次20秒,流速3 ml/20s。
  • LAI参数:Er:YAG激光,P400FL工作尖置于根管口,25 pps,50 mJ,300 μs。
  • SAI-E和SAI-EA使用25/.04工作尖,MA使用25/.06牙胶尖。

临床怎么用

  • 对于复杂根管系统的清洁,可优先考虑激光激活荡洗(LAI)以获得最佳整体效果。
  • 在峡部区域,EDDY或传统针头冲洗可作为LAI的替代方案。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: This study aimed to evaluate the removal of a biofilm-mimicking hydrogel from isthmus structures in a simulated complex root canal system consisting of 2 curved root canals by Laser-activated irrigation (LAI, AdvErl Evo, Morita) and mechanical activation techniques. METHODS: A 3D-printed root canal model with 2 parallel root canals (60°-curvature, radius 5 mm, dimension 25/.06) with a total length of 20 mm connected via isthmuses (2.5 × 0.4 × 0.2 mm) at 5 mm and 8 mm from the apical endpoint and with lateral canals (diameter 0.2 mm) in all directions at 2, 5, and 8 mm from the apex was filled with a colored biofilm-mimicking hydrogel. Irrigation protocols under continuous irrigation with distilled water (3 × 20s per root canal; 3 ml/20s; n = 20) included conventional needle irrigation (=NI); manual agitation (=MA, gutta-percha point 25/.06); EndoActivator (=SAI-EA, 25/.04); EDDY (=SAI-E, 25/.04); ultrasonically-activated irrigation (=UAI) and LAI (Er:YAG-laser; P400FL tip at canal entrance; 25pps, 50 mJ, 300μs). Removal of the hydrogel was determined as a percentage via standardized photos through a microscope. Statistical analysis was performed using Kruskal-Wallis and Conover tests (P = .05). RESULTS: Laser-activated irrigation (LAI) was associated with the greatest removal of hydrogel from the entire root canal system (P < .05), followed by SAI-E. No significant differences were reported for the coronal isthmus between LAI, SAI-E, NI, and MA (P > .05), but inferior results for SAI-EA and UAI (P < .05). In the apical isthmus, all techniques outperformed UAI (P < .05), with LAI, SAI-E, and NI showing the best results (P < .05). CONCLUSIONS: Laser-activated irrigation (LAI) was superior to other irrigation techniques in the entire root canal system. SAI-E and NI performed comparable to LAI in the isthmuses.

来源:J Endod,查看原文。版权归原作者所有。

引用

  1. Donnermeyer D, Dust PC, Schäfer E, et al. Comparative Analysis of Irrigation Techniques for Cleaning Efficiency in Isthmus Structures. J Endod. 2024. PMID: 38382735. doi:10.1016/j.joen.2024.02.008

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。