智能手机生成三维面部图像能否用于常规口腔临床?一项验证研究
结果显示三种方法均值无统计学显著差异,DA与SGI、3dMD与SGI分别有67%和74%的测量达到临床可接受差异(≤3 mm或≤5°),总体系统偏倚约±0.2 mm,SGI在面部中央及平坦区域更准确,但整体面部精度有限。
要点
- SGI与DA、3dMD的均值无统计学显著差异,总体系统偏倚约±0.2 mm。
- DA与SGI、3dMD与SGI分别有67%和74%的测量差异在≤3 mm或≤5°内。
- SGI适用于面部中央及平坦区域的宏观比例评估或患者教育。
核心结论
- SGI相对DA和3dMD表现出尚可的真实性,面部中央区域和平坦区域更准确。
- SGI整体面部精度有限,临床应用受限,从临床角度看更理想的是全脸精度。
研究设计与样本
- 验证性研究,纳入22例正颌手术患者(11男11女,平均年龄26.2 ± 5.3岁)。
- 直接在每个标记面部及三维图像上测量42个标志点间距,比较SGI与DA、3dMD的测量值。
主要发现
- SGI测量值略高于DA和3dMD,但三种方法标志点间测量均值无统计学显著差异。
- DA与SGI之间67%的测量、3dMD与SGI之间74%的测量达到临床可接受差异(≤3 mm或≤5°),一致性良好。
- 三种方法间总体系统偏倚约±0.2 mm。
- DA与SGI的平均绝对差异在线性测量中最高为1.41 ± 0.33 mm,在角度测量中最高为3.07 ± 0.73°。
临床怎么用
- SGI可用于涉及面部中央和平坦区域的区域特异性宏观比例面部评估。
- SGI可用于患者教育目的,前提是不要求精度在3 mm和5°以内。
局限性
- SGI整体面部精度有限,临床应用受限。
- 从临床应用角度看,SGI的全脸精度更理想。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: To compare the accuracy of smartphone-generated three-dimensional (3D) facial images to that of direct anthropometry (DA) and 3dMD with the aim of assessing the validity and reliability of smartphone-generated 3D facial images for routine clinical applications. MATERIALS AND METHODS: Twenty-five anthropometric soft-tissue facial landmarks were labelled manually on 22 orthognathic surgery patients (11 males and 11 females; mean age 26.2 ± 5.3 years). For each labelled face, two imaging operations were performed using two different surface imaging systems: 3dMDface and Bellus3D FaceApp. Next, 42 inter-landmark facial measurements amongst the identified facial landmarks were measured directly on each labelled face and also digitally on 3D facial images. The measurements obtained from smartphone-generated 3D facial images (SGI) were statistically compared with those from DA and 3dMD. RESULTS: SGI had slightly higher measurement values than DA and 3dMD, but there was no statistically significant difference between the mean values of inter-landmark measures across the three methods. Clinically acceptable differences (≤3 mm or ≤5°) were observed for 67 % and 74 % of measurements with good agreement between DA and SGI, and 3dMD and SGI, respectively. An overall small systematic bias of ± 0.2 mm was observed between the three methods. Furthermore, the mean absolute difference between DA and SGI methods was highest for linear (1.41 ± 0.33 mm) as well as angular measurements (3.07 ± 0.73°). CONCLUSIONS: SGI demonstrated fair trueness compared to DA and 3dMD. The central region and flat areas of the face in SGI are more accurate. Despite this, SGI have limited clinical application, and the panfacial accuracy of the SGI would be more desirable from a clinical application standpoint. CLINICAL SIGNIFICANCE: The usage of SGI in clinical practice for region-specific macro-proportional facial assessment involving central and flat regions of the face or for patient education purposes, which does not require accuracy within 3 mm and 5° can be considered.
引用
- Singh P, Hsung RT, Ajmera DH, et al. Can smartphones be used for routine dental clinical application? A validation study for using smartphone-generated 3D facial images. J Dent. 2023. PMID: 37944629. doi:10.1016/j.jdent.2023.104775
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。