自主牙种植机器人系统骨切开时骨内温度变化与钻削冲量数据的相关性:一项体外研究
本研究在体外牛肋模型上使用自主牙种植机器人系统(ADIR),通过红外相机和六轴力/力矩传感器分别测量骨内温度和钻削冲量数据,共测试800个不同参数钻头并另设200个独立测试集。结果发现钻削冲量数据与骨内温度变化高度线性相关,基于XGboost的机器学习预测模型在测试集中误差最小。
要点
- 钻削冲量数据与骨内温度变化高度线性相关。
- 较低钻速、较小钻径、更严重磨损和较厚皮质骨与更高骨内温度变化相关。
- XGboost模型在四种机器学习模型中预测误差最小。
核心结论
- 骨切开过程中钻削冲量数据与骨内温度变化密切相关。
- 机器学习预测模型可能有助于预防热骨损伤和机器人辅助种植手术的智能设计。
研究设计与样本
- 体外牛肋模型结合自主牙种植机器人系统(ADIR)进行骨切开实验。
- 使用红外相机和六轴力/力矩传感器分别测量骨内温度和钻削冲量数据。
- 共测试800个不同参数(钻径、钻头磨损、钻速、皮质骨厚度)的钻头,并设200个独立测试集。
主要发现
- 较低钻速、较小钻径、更严重磨损和较厚皮质骨与更高骨内温度变化、更长暴露时间及钻削冲量数据改变相关。
- Pearson相关分析显示钻削冲量数据与热变化高度线性相关。
- 四种机器学习模型中XGboost模型在测试集误差测量最小。
临床怎么用
- 钻削冲量数据可用于实时预测骨内温度变化。
- XGboost模型可作为机器人辅助种植手术中预防热骨损伤的参考工具。
局限性
- 本研究为体外概念验证研究,结果需进一步临床验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: This study aims at examining the correlation of intraosseous temperature change with drilling impulse data during osteotomy and establishing real-time temperature prediction models. MATERIALS AND METHODS: A combination of in vitro bovine rib model and Autonomous Dental Implant Robotic System (ADIR) was set up, in which intraosseous temperature and drilling impulse data were measured using an infrared camera and a six-axis force/torque sensor respectively. A total of 800 drills with different parameters (e.g., drill diameter, drill wear, drilling speed, and thickness of cortical bone) were experimented, along with an independent test set of 200 drills. Pearson correlation analysis was done for linear relationship. Four machining learning (ML) algorithms (e.g., support vector regression [SVR], ridge regression [RR], extreme gradient boosting [XGboost], and artificial neural network [ANN]) were run for building prediction models. RESULTS: By incorporating different parameters, it was found that lower drilling speed, smaller drill diameter, more severe wear, and thicker cortical bone were associated with higher intraosseous temperature changes and longer time exposure and were accompanied with alterations in drilling impulse data. Pearson correlation analysis further identified highly linear correlation between drilling impulse data and thermal changes. Finally, four ML prediction models were established, among which XGboost model showed the best performance with the minimum error measurements in test set. CONCLUSION: The proof-of-concept study highlighted close correlation of drilling impulse data with intraosseous temperature change during osteotomy. The ML prediction models may inspire future improvement on prevention of thermal bone injury and intelligent design of robot-assisted implant surgery.
引用
- Zhao R, Xie R, Ren N, et al. Correlation between intraosseous thermal change and drilling impulse data during osteotomy within autonomous dental implant robotic system: An in vitro study. Clin Oral Implants Res. 2023. PMID: 38031528. doi:10.1111/clr.14222
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。