自主机器人种植手术在全口无牙颌患者中的精度:回顾性病例系列研究
本研究评估自主机器人计算机辅助种植手术(r-CAIS)在全口无牙颌患者中的种植体植入精度。全局冠部、全局根尖部和角度偏差分别为0.67±0.37 mm(95%CI:0.58-0.77 mm)、0.69±0.37 mm(95%CI:0.59-0.78 mm)和1.27°±0.59°(95%CI:1.42°-1.11°)。结论认为该技术精度较高,但需进一步试验提供更多硬临床证据。
要点
- 自主机器人种植手术在全口无牙颌患者中精度较高,角度偏差控制良好。
- 全局冠部偏差0.67±0.37 mm,全局根尖部偏差0.69±0.37 mm。
核心结论
- 自主机器人计算机辅助种植手术(r-CAIS)在全口无牙颌患者中精度较高,尤其在角度偏差控制方面表现良好。
- 该技术有望成为治疗全口无牙颌患者的精准技术,但仍需进一步试验提供更多硬临床证据。
研究设计与样本
- 回顾性病例系列研究,纳入全口无牙颌及终末牙列患者接受r-CAIS技术。
- 术中由自主机器人系统在术者监督下自动完成种植体窝洞制备和植入,术后CBCT扫描评估计划与植入种植体的偏差。
主要发现
- 全局冠部偏差为0.67±0.37 mm(95%CI:0.58-0.77 mm)。
- 全局根尖部偏差为0.69±0.37 mm(95%CI:0.59-0.78 mm)。
- 角度偏差为1.27°±0.59°(95%CI:1.42°-1.11°)。
临床怎么用
- 对于全口无牙颌患者,可考虑采用自主机器人计算机辅助种植手术以实现较高的植入精度。
- 术前需基于CBCT制作个性化定位标记和手术计划,术中由术者监督机器人自动完成窝洞制备和植入。
局限性
- 结论指出仍需进一步试验提供更多硬临床证据。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: This study evaluated the accuracy of dental implant placement using the robotic computer-assisted implant surgery (r-CAIS) technology in fully edentulous patients. MATERIALS AND METHODS: Fully edentulous and terminal dentition patients were enrolled for r-CAIS technology. Based on the cone-beam computed tomography (CBCT) examination, a customized positioning marker and a preoperative surgical plan were created before surgery. During the implant surgery, the implant osteotomy and placement were automatically performed using an autonomous robotic surgery system under the surgeon's supervision. A postoperative CBCT scan was used to determine the discrepancies between the planned and placed implants. RESULTS: Ten patients with 59 implants underwent autonomous robotic surgery. No adverse surgical events occurred. The deviations of global coronal, global apical, and angular were 0.67 ± 0.37 mm (95% CI: 0.58-0.77 mm), 0.69 ± 0.37 mm (95% CI: 0.59-0.78 mm), and 1.27° ± 0.59° (95% CI: 1.42°-1.11°), respectively. CONCLUSIONS: The autonomous r-CAIS technology proved an accurate surgical approach for implant placement in fully edentulous patients due to the control of the angular deviation. Autonomous robotic surgery seems promising as an accurate technology for treating fully edentulous patients. However, further trials are required to provide more hard clinical evidence.
引用
- Li P, Chen J, Li A, et al. Accuracy of autonomous robotic surgery for dental implant placement in fully edentulous patients: A retrospective case series study. Clin Oral Implants Res. 2023. PMID: 37855331. doi:10.1111/clr.14188
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。