文献速递口腔修复与种植

新型种植系统与简化钻孔方案在迷你猪口腔种植模型中的性能研究

Performance of a new implant system and drilling protocol-A minipig intraoral dental implant model study.
PubMed 文献 · Clin Oral Implants Res · 2023-10-19

本研究采用八只尤卡坦迷你猪口腔种植模型,评估新型三卵圆形截面种植体及低速无冲洗简化备洞方案的安全性与有效性。结果显示试验组骨-种植体接触率(BIC)显著高于对照组(74.1% vs. 60.9%,p < .001),支持该新系统安全有效,但需人体研究证实。

要点

  • 试验组组织学BIC为74.1%,显著高于对照组60.9%(p < .001)。
  • 试验组平台水平炎症轻度但显著高于对照组,其余指标无显著差异。
  • MicroCT显示试验组BIC与骨小梁厚度显著更高,舌侧首BIC距离显著更大。

核心结论

  • 本研究结果支持新型种植系统及简化备洞方案的安全性与有效性。
  • 尚需开展人体研究以证实上述发现。

研究设计与样本

  • 采用八只尤卡坦迷你猪口腔种植模型。
  • 对照组为2枚Ø3.5 × 10 mm、背锥形颈圈与圆形截面种植体;试验组为2枚Ø3.5 × 11 mm、三卵圆形颈圈与截面种植体。

主要发现

  • 组织学BIC试验组高于对照组(74.1% vs. 60.9%,p < .001)。
  • 平台水平炎症试验组显著高于对照组,但程度轻微。
  • MicroCT显示试验组BIC与骨小梁厚度显著高于对照组。
  • 舌侧位点试验组首BIC距离显著大于对照组。
  • 两组间未观察到其他显著差异。

临床怎么用

  • 新型三卵圆形截面种植体配合低速无冲洗简化备洞方案在动物模型中表现安全有效。
  • 临床转化前仍需人体研究验证其性能。

局限性

  • 研究采用迷你猪动物模型,结论外推至人体需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIM: A new implant system encompassing implants with a tri-oval cross-sectional design and a simplified site preparation protocol at low speed and no irrigation has been developed. The objective of this study was to assess the safety and efficacy of the new implant system using the minipig intraoral dental implant model. METHODS: Eight Yucatan minipigs were included. Twelve weeks after extractions, four implants per animal were randomly placed and allowed to heal transmucosal for 13 weeks: two Ø3.5 × 10 mm implants with a back-tapered collar and circular cross-section (control) and two Ø3.5 × 11 mm implants with tri-oval collar and cross-section (test). MicroCT and histological analysis was performed. RESULTS: Thirty-two implants were placed; one implant for the control group was lost. Histologically, BIC was higher in the test compared with the control group (74.1% vs. 60.9%, p < .001). At the platform level, inflammation was statistically significantly higher albeit mild in the test compared with the control group. No other significant differences were observed between groups. MicroCT analysis showed that bone-to-implant-contact (BIC) and trabecular thickness were statistically significantly higher for the test than the control group. Test group had significantly higher first BIC distance than controls on lingual sites. CONCLUSIONS: The present study results support the safety and efficacy of the new dental implant system and simplified site preparation protocol; human studies should be carried out to confirm these findings.

来源:Clin Oral Implants Res,查看原文。版权归原作者所有。

引用

  1. Musskopf ML, Finger Stadler A, Fiorini T, et al. Performance of a new implant system and drilling protocol-A minipig intraoral dental implant model study. Clin Oral Implants Res. 2023. PMID: 37855174. doi:10.1111/clr.14194

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。