文献速递牙体牙髓病

新型牙髓活力测试仪与常用方法的效果与安全性比较:交叉研究

Comparison of the Efficacy and Safety of Pulp Sensitivity Testing Results Using a New Pulp Testing Device versus Some Commonly Used Methods: A Crossover Study.
PubMed 文献 · J Endod · 2023-10-14

结果显示PSTT在热测与透照上高效安全且总体测试时间更短,但冷测与电测的响应比例低于GS方法,需改进。

要点

  • PSTT热测与透照高效安全,总体测试时间显著短于GS方法。
  • PSTT冷测与电测响应比例低于GS方法,需改进刺激效能。
  • 未观察到组织损伤,安全性可接受。

核心结论

  • PSTT可高效、安全地提供热测与透照用于牙髓诊断,但冷测与电测效能需改进。

研究设计与样本

  • 结果用IBM SPSS分析,P < .05。

主要发现

  • GS方法冷测与电测响应牙比例高于PSTT(99.4% vs 93.1%;99.6% vs 97.3%)。
  • GS方法热测响应牙比例低于PSTT(50.0% vs 68.1%)。
  • 温度与电测敏感性比例差异有统计学意义(P < .001 与 P = .004),透照无差异。
  • GS方法冷测时间显著更短(P = .024);PSTT热测(P < .001)、电测(P = .048)、透照(P = .001)时间更短。
  • PSTT总体测试时间显著短于GS(P = .03),未观察到组织损伤。

临床怎么用

  • 可将PSTT用于热测与透照辅助牙髓诊断,并利用其总体时间较短的优势。
  • 冷测与电测结果解读需谨慎,必要时结合GS方法验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: The efficacy and safety of a single novel electronic pulp sensitivity tester with a transilluminator (PSTT) capable of providing 4 different tests was compared with gold standard (GS) pulp testing methods. METHODS: Four hundred eighty teeth, including 3 from each quadrant and their contralateral and opposing teeth, were randomly assessed using the PSTT or GS methods. Seven days later, the same teeth were assessed using the method not used earlier so that all teeth were evaluated using both methods. Sixty previously root canal-treated teeth, serving as negative controls, were assessed identically to the experimental groups. Results were analyzed using IBM SPSS software (IBM Corp, Armonk, NY) (P < .05). RESULTS: A higher percentage of teeth responded to cold and electricity using GS methods compared with the PSTT (99.4 vs 93.1 and 99.6 vs 97.3, respectively). A lower percentage of teeth responded to heat using the GS method compared with the PSTT (50.0 vs 68.1). Chi-square tests determined differences in the proportion of sensitivity to temperature, and electric testing methods were statistically significant (P < .001 and P = .004, respectively). No difference in the proportion of sensitivity was observed for transillumination. The Wilcoxon signed rank test determined significantly shorter cold testing times using the GS method (P = .024). Shorter testing times were observed using the PSTT for heat (P < .001), electric pulp testing (P = .048), and transillumination (P = .001). The overall PSTT testing time was significantly shorter than the GS testing time (P = .03). Tissue injury was not observed. CONCLUSIONS: The PSTT efficiently and safely provided heat and transillumination for pulpal diagnosis. Improvements are needed to enhance the cold and electric stimulus efficacy.

来源:J Endod,查看原文。版权归原作者所有。

引用

  1. Tordik PA, Torabinejad M, Fevelo CM, et al. Comparison of the Efficacy and Safety of Pulp Sensitivity Testing Results Using a New Pulp Testing Device versus Some Commonly Used Methods: A Crossover Study. J Endod. 2023. PMID: 37839772. doi:10.1016/j.joen.2023.10.003

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。