新型预制辅助装置附着于扫描体对全牙弓多种植体口内扫描精度的影响:体外研究
本研究通过体外模型评估新型预制辅助装置(扫描体卡扣,SBC)的不同几何特征和位置对全牙弓多种植体口内扫描精度的影响。结果显示,附着SBC可提升扫描正确度,其中靠近黏膜的SBC正确度更优,平面形态的SBC精度更佳。结论提示SBC可提高全牙弓种植扫描精度。
要点
- SBC可提升全牙弓多种植体口内扫描的正确度。
- 靠近黏膜的SBC正确度更优,平面形态SBC精度更佳。
- 低位置平面SBC可能是优化长跨度扫描精度的关键。
核心结论
- 附着SBC的不同位置和几何特征显著影响全牙弓多种植体口内扫描精度。
- 靠近黏膜的SBC可获得更优正确度,平面形态有利于精度。
- SBC可提高全牙弓种植扫描精度。
研究设计与样本
- 体外研究,使用3D打印无牙上颌4种植体模型和不同几何特征(平面或曲面)的SBC。
- 每组(CO、LC、HC、LF、HF)进行20次扫描,参考扫描由实验室扫描仪数字化A组模型获得。
- 使用单因素方差分析、独立样本t检验、Kruskal-Wallis检验和Mann-Whitney检验,显著性水平α=0.05。
主要发现
- 带SBC的组别正确度提升,其中LF组正确度最佳。
- LF与HC(p < .01)、LF与HF(p < .001)、LC与HF(p < .01)之间正确度差异显著。
- LF和HF组精度提升,LF精度最佳,优于LC(p < .001)和HC(p < .001)。
- HF精度优于LC(p < .001)和HC(p < .001)。
临床怎么用
- SBC可用于提升全牙弓种植口内扫描精度。
- 低位置和平面形态的人工标记可能是优化长跨度扫描精度的关键要素。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: To examine the effect of novel prefabricated auxiliary devices with different geometric features called Scan Body Clasp (SBC) at different levels on the accuracy of intraoral scanning of complete-arch with multiple implants. METHODS: An edentulous maxilla 4-implant model and SBCs with different geometric features (flat or curved) were fabricated by a 3D printer (AccuFab-C1s, 3DShining, Hangzhou, China). Test scans were performed using an intraoral scanner (Aoralscan 3, 3DShining, Hangzhou, China) software version 1.0.0.3104 under different scenarios: group A (CO), without any SBCs; group B&C (LC&HC), with curved SBCs adjacent to and away from the mucosa; group D&E (LF&HF), with flat SBCs adjacent to and away from the mucosa. 20 scans were done for each group (CO, LC, HC, LF and HF). Reference Scans were obtained by digitizing the model in group A using a dental laboratory scanner (D2000, 3Shape, Copenhagen, Denmark). The related files were imported into inspection software for trueness and precision assessment. Statistical analysis was performed with One-way ANOVA, Independent-Sample T test for trueness values. Kruskal-Wallis test and Mann-Whitney test were used to assess the precision values. The level of significance was set at α=0.05. RESULTS: Groups with SBCs demonstrated trueness enhancement, among which LF revealed the best trueness. Significant differences were also found between LF and HC (p < .01), LF and HF (p < .001), LC and HF (p < .01). LF and HF showed precision enhancement. The best precision was LF, which was found to be more precise than LC (p < .001) and HC (p < .001). HF was more precise than LC (p < .001) and HC (p < .001). CONCLUSIONS: Attaching the scan bodies with SBCs at different levels significantly influenced the scanning accuracy. The SBCs near the mucosa result in superior trueness, while the flat morphology benefits the precision. CLINICAL SIGNIFICANCE: The results demonstrated the feasibility of the SBCs in enhancing intraoral complete-arch implant scanning accuracy. Among the configurations tested in the present study, low-level and flat surfaces of the artificial landmarks may be the potential pivotal elements to optimizing long-span scanning accuracy.
引用
- Wu HK, Wang J, Chen G, et al. Effect of novel prefabricated auxiliary devices attaching to scan bodies on the accuracy of intraoral scanning of complete-arch with multiple implants: An in-vitro study. J Dent. 2023. PMID: 37714453. doi:10.1016/j.jdent.2023.104702
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。