文献速递牙体牙髓病

锥形束CT与电子根尖定位仪检测模拟根管穿孔准确性的体外比较

In Vitro Comparison of the Accuracy of Cone-Beam Computed Tomography Scanning and Electronic Apex Locators in Detection of Simulated Root Perforations in Different Localizations.
PubMed 文献 · J Endod · 2023-09-21

结果显示根尖三分之一组各设备均较成功,根中三分之一组所有设备成功率下降,其中Propex II显著优于CBCT。

要点

  • 根尖三分之一穿孔时,电子根尖定位仪与CBCT均可较准确判定穿孔区域。
  • 根中三分之一穿孔时所有设备成功率下降,Propex II准确性显著优于CBCT。
  • CBCT对根尖三分之一穿孔的判定准确性高于根中三分之一穿孔。

核心结论

  • 根尖三分之一穿孔时电子根尖定位仪与CBCT均能成功确定穿孔区域。
  • 根中三分之一穿孔时本研究中所有设备的成功率均下降。

研究设计与样本

  • 按穿孔位置分为根尖三分之一组与根中三分之一组。
  • 以立体显微镜下测得咬合边缘至穿孔起始的距离为实际工作长度,与两种电子根尖定位仪及CBCT测得的工作长度比较差值。

主要发现

  • 根尖三分之一组中Dentaport ZX与CBCT成像之间差异有统计学意义。
  • 根尖三分之一组中Dentaport ZX与Propex II、Propex II与CBCT之间差异无统计学意义。
  • 根中三分之一组中Propex II显著优于CBCT成像。
  • Propex II在根尖三分之一组与根中三分之一组检测穿孔区域的准确性之间无统计学意义。
  • CBCT在根尖三分之一组检测穿孔区域比根中三分之一组更准确。

临床怎么用

  • 根尖三分之一可疑穿孔可考虑电子根尖定位仪或CBCT辅助判定。
  • 根中三分之一穿孔时设备准确性下降,解读结果需谨慎。

局限性

  • 为体外模拟穿孔研究,穿孔直径固定为1 mm,结果外推临床需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: This study aimed to compare the accuracy of cone-beam computed tomographic (CBCT) scanning and 2 different electronic apex locators (EALs) in the detection of simulated root canal perforations in different localizations. METHODS: Eighty human mandibular premolars were prepared and divided into 2 groups according to the localization of the perforation area, having a 1-mm diameter. The distance between the occlusal edge and the beginning of the perforation was measured under a stereomicroscope (actual working length) using 2 EALs (Propex II [Dentsply Maillefer, Ballaigues, Switzerland] and Dentaport ZX [J. Morita, Tokyo, Japan]) (electronic working length) and CBCT images (CBCT working length). The calculations were made by subtracting the actual working length from the electronic working length and the CBCT working length to determine the differences. RESULTS: In the apical third perforation group, there was a significant difference between Dentaport ZX and CBCT imaging. There was no significant difference between Dentaport ZX and Propex II and Propex II and CBCT measurements. In the middle third perforation group, Propex II was significantly more successful than CBCT imaging. In addition, there was no significance between the accuracy of Propex II in detecting the perforation area between the apical third perforation group and the middle third perforation group. CBCT imaging was more accurate in detecting the perforation area in the apical third perforation group than in the middle third perforation group. CONCLUSIONS: In cases in which the root canal perforation was in the apical third of the root, both the EAL and CBCT imaging were successful in determining the perforation area, whereas the success rate of all devices used in this study decreased when the perforation area was in the middle third of the root.

来源:J Endod,查看原文。版权归原作者所有。

引用

  1. Harorlı H, Koç S, Kuştarcı A, et al. In Vitro Comparison of the Accuracy of Cone-Beam Computed Tomography Scanning and Electronic Apex Locators in Detection of Simulated Root Perforations in Different Localizations. J Endod. 2023. PMID: 37739340. doi:10.1016/j.joen.2023.09.006

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。