13种机械预备系统在模拟根管中应力分布的准三维动态光弹性分析
各系统均有应力集中,大锥度、低转速、往复运动及热处理镍钛器械应力较高,冠、中、根尖三分之一处组间差异显著(p<.05)。
要点
- 所有预备系统均非无应力,应力水平和分布存在差异。
- 大锥度、低转速、往复运动及热处理镍钛器械应力较高。
- 冠、中、根尖三分之一处各组应力评分差异显著。
核心结论
- 无任何根管预备方案无应力,不同系统应力集中程度不同。
- 大锥度、低转速、往复运动及热处理镍钛合金器械应力分布较高。
- 多种因素影响应力大小及在根管壁上的分布模式。
研究设计与样本
- 按13种预备系统分组,每组5个样本。
- 用圆偏振光暗场装置和数码相机记录整个预备过程。
- 提取器械到达冠、中、根尖三分之一时的视频帧,由两名独立观察者用0-5半定量评分评估应力。
主要发现
- 观察者间和观察者内一致性分别为0.98和1。
- 多数器械在三个三分之一处评分≤2,表现可接受。
- HyFlex EDM 25.12在冠和中三分之一、Reciproc Blue R25和Reciproc R25在冠和根尖三分之一、R-Motion 30.04和VDW.ROTATE 20.05在根尖三分之一评分高于3。
- 统计显示冠、中、根尖三分之一处各系统间差异显著(p<.05)。
关键数值与推荐
- 应力半定量评分采用0-5分制。
- 观察者间一致性Cohen's Kappa为0.98,观察者内为1。
- 统计比较用Kruskal-Wallis检验及Bonferroni校正事后两两比较,α=5%。
临床怎么用
- 选择预备系统时需考虑其应力分布特征,尤其在大锥度、低转速或往复运动器械。
- 对弯曲根管,关注冠、中、根尖三分之一的应力集中差异。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIM: The aim of this study is to compare the stress produced on the internal walls of simulated canals by nine rotary and four reciprocating systems. METHODOLOGY: Sixty-five isotropic transparent blocks containing a 60° curved and tapered simulated canal were selected and distributed into 13 groups (n = 5) according to the preparation system: BioRace, HyFlex EDM, iRaCe, Mtwo, One RECI, ProTaper Next, RaCe EVO, Reciproc, Reciproc Blue, R-Motion, VDW.ROTATE, XP-Endo Rise Shaper, and XP-Endo Shaper. Each resin block was mounted in a vice and a digital camera recorded the entire sequence of each preparation system through a circular polariscope set for dark field analysis. The video frames when each instrument reached the end of the coronal, middle, and apical thirds of the canal were extracted from the recordings and analysed by two independent observers regarding the stress generated on the canal walls using a semi-quantitative evaluation on a 0-5 scale. Intra- and inter-observer agreement were subjected to the Cohen's Kappa coefficient test, whilst the experimental results were compared using Kruskal-Wallis test post hoc pairwise comparisons with Bonferroni correction (α = 5%). RESULTS: The inter- and intra-observer agreement were 0.98 and 1, respectively. Most instruments demonstrated acceptable performance (scores ≤ 2) in all thirds. Other instruments, such as the HyFlex EDM 25.12 (coronal and middle thirds), Reciproc Blue R25 and Reciproc R25 (coronal and apical thirds), R-Motion 30.04 (apical third), and VDW.ROTATE 20.05 (apical third) showed scores higher than 3. Statistical analysis revealed a significant difference amongst the tested systems at the coronal, middle, and apical thirds (p < .05). CONCLUSION: None of the canal instrumentation protocols were stress-free, showing varying levels of stress concentrations. Various factors seemed to influence the magnitude of stress and its distribution pattern on the canal walls. Overall, instruments characterized by a larger taper, lower speed, reciprocating motion, and made of heat-treated NiTi alloy exhibited higher patterns of stress distribution.
引用
- Moriya PT, de Carvalho KKT, Kishen A, et al. Quasi-3D dynamic photoelastic analysis of stress distribution during preparation of simulated canals with 13 mechanical preparation systems. Int Endod J. 2023. PMID: 37566206. doi:10.1111/iej.13961
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。