种植体成形术与窄种植体骨折风险:一项实验室研究

Implantoplasty and the risk of fracture of narrow implants with advanced bone loss: A laboratory study.
PubMed 文献 · Clin Oral Implants Res · 2023-07-18

本研究评估种植体成形术对模拟严重骨吸收条件下不同型号和材料窄种植体最大失败强度的影响。结果显示,种植体成形术显著降低窄种植体最大失败强度,组织水平种植体较骨水平种植体受影响更严重,且该负面影响与种植体材料无关。

要点

  • 种植体成形术显著降低窄种植体最大失败强度(p < .002)。
  • 组织水平种植体经种植体成形术后动态加载骨折率显著增加(p = .001)。
  • 窄单颗组织水平种植体伴严重水平骨吸收时,种植体成形术增加正常功能下的骨折风险。

核心结论

  • 基于单一制造商种植体数据,种植体成形术对伴严重种植体周围炎的窄种植体骨折强度有显著负面影响。
  • 组织水平种植体较骨水平种植体受影响更严重,正常咀嚼力下失败风险增加。
  • 种植体成形术对组织水平种植体的负面影响与种植体材料(Ti或TiZr)无关。

研究设计与样本

  • 实验室研究,使用单一制造商窄平行壁种植体(直径3.3 mm×长10 mm),内连接。
  • 比较不同设计[骨水平(BL)、组织水平(TL)]和材料[钛IV级(Ti)、钛锆合金(TiZr)]。
  • 一半种植体冠状5 mm行种植体成形术,其余为对照,每组7颗种植体。
  • 包含动态加载后进行最大载荷强度测试。

主要发现

  • 动态加载期间,骨水平种植体骨折率低且与种植体成形术无关,组织水平种植体经种植体成形术后骨折率较对照显著增加(p = .001)。
  • 种植体成形术导致最大失败强度降低百分比为1.3%-25.4%,TiZr骨水平种植体受影响最小。
  • 行种植体成形术的种植体较未行者、组织水平种植体较骨水平种植体最大失败强度显著更低(p < .002)。
  • 种植体材料对最大失败强度无显著影响(p = .845)。

临床怎么用

  • 窄单颗组织水平种植体伴严重水平骨吸收(如5 mm)时,行种植体成形术需警惕正常功能下的骨折风险。

局限性

  • 数据仅来自单一制造商种植体。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: To assess the impact of implantoplasty (IP) on maximum implant failure strength of narrow diameter implants of different type/design and material, with simulated advanced bone loss. MATERIALS AND METHODS: Narrow, parallel-walled implants (3.3 mm in diameter × 10 mm long) with an internal connection of different type/design [bone level (BL), tissue level (TL)] and material [Titanium grade IV (Ti), Titanium-Zirconium alloy (TiZr)] from one specific manufacturer were used. Half of the implants were subjected to IP in their coronal 5 mm; the remaining were used as controls (seven implants per group). Dynamic loading prior to maximum load strength testing was included. RESULTS: During dynamic loading, the fracture rate of BL implants was low and independent of IP, while that of TL implants increased significantly with IP compared with controls (p = .001). Maximum implant failure strength reduction (in %) due to IP, was 1.3%-25.4%; TiZr BL implants were least affected. Implants subjected to IP compared to those without IP as well as TL implants compared to BL implants showed a significantly lower maximum implant failure strength (p < .002); implant material was not significant (p = .845). CONCLUSIONS: Based on data from implants of one specific manufacturer, IP has a significant negative impact on the fracture strength of narrow implants suffering from advanced peri-implantitis. TL implants have been more severely affected compared to BL implants and presented an increased risk for failure during normal chewing forces. In addition, this negative impact of IP on TL implants was independent of the implant material (i.e., Ti or TiZr). CLINICAL RELEVANCE: Narrow single TL implants with advanced horizontal bone loss (e.g., 5 mm), when subjected to IP, appear to have an increased fracture risk during normal function.

来源:Clin Oral Implants Res,查看原文。版权归原作者所有。

引用

  1. Stavropoulos A, Bertl K, Isidor F, et al. Implantoplasty and the risk of fracture of narrow implants with advanced bone loss: A laboratory study. Clin Oral Implants Res. 2023. PMID: 37464268. doi:10.1111/clr.14132

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。