动态导航系统与三维显微镜取出分离根管锉的体外对比研究
结果显示显微镜组成功率较高但差异无统计学意义(P > .05),而显微镜组在根管偏移比例、取出时间和根管体积变化方面均显著更优(P < .05)。结论认为动态导航环钻取出虽可行,但在治疗时间、操作误差和体积变化方面逊于三维显微镜引导。
要点
- 三维显微镜引导取出分离器械的成功率高于动态导航,但差异无统计学意义。
- 显微镜引导的根管偏移比例显著更低,取出时间更短,根管体积变化更小。
- 动态导航环钻配合取出器可取出分离器械,但操作误差和体积损失更大。
核心结论
- 动态导航引导环钻配合取出器可取出分离器械,但在治疗时间、操作误差和根管体积变化方面逊于三维显微镜引导。
研究设计与样本
- 标准化开髓、建立滑行通道并用K锉(10号和15号)保持通畅后,将牙排列于3D模型上(每象限3颗,每模型6颗)。
- 使用距尖端4 mm处开槽并在根尖三分之一处折断的控忆热处理镍钛旋转锉(#25/0.04),在两种引导下用C-FR1 Endo取出系统取出碎片。
- 测量成功率、根管偏移、治疗时间和体积变化,使用IBM SPSS软件以0.05为显著性水平进行统计分析。
主要发现
- 显微镜引导组成功率高于动态导航系统引导组,但差异无统计学意义(P > .05)。
- 显微镜引导钻的根管偏移比例显著更低(P < .05)。
- 显微镜引导取出碎片的时间显著更短(P < .05)。
- 显微镜引导的根管体积变化显著更小(P < .05)。
临床怎么用
- 在取出根尖三分之一分离的旋转镍钛锉时,优先考虑三维显微镜引导以减少根管偏移和体积损失。
- 动态导航可作为替代方案,但需注意其治疗时间更长、操作误差和体积变化更大。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: This study aimed to compare the effectiveness of a dynamic navigation system and a three-dimensional microscope in retrieving broken rotary Nickel-Titanium files when using trepan burs and the extractor system. MATERIALS AND METHODS: Thirty maxillary first bicuspids with 60 separate roots were split into 2 comparable groups based on a comprehensive cone beam computed tomography analysis of the root length and curvature. After standardized access opening, glide paths, and patency attainment with the K file (sizes 10 and 15), the teeth were arranged on 3D models (three per quadrant, six per model). Subsequently, controlled-memory heat-treated Nickel-Titanium rotary files (#25/0.04) were notched 4 mm from the tips and fractured at the apical third of the roots. The C-FR1 Endo file removal system was employed under both guidance to retrieve the fragments, and the success rate, canal aberration, treatment time, and volumetric changes were measured. The statistical analysis was performed using IBM SPSS software at a significance level of 0.05. RESULTS: The microscope-guided group had a higher success rate than the dynamic navigation system guidance, but the difference was insignificant (P > .05). In addition, the microscope-guided drills resulted in a substantially lower proportion of canal aberration, shorter time to retrieve the fragments and less change in the root canal volume (P < .05). CONCLUSION: Although dynamically guided trephining with the extractor can retrieve separated instruments, it is inferior to three-dimensional microscope guidance regarding treatment time, procedural errors, and volume change.
引用
- Karim MH, Faraj BM. Comparative Evaluation of a Dynamic Navigation System versus a Three-dimensional Microscope in Retrieving Separated Endodontic Files: An In Vitro Study. J Endod. 2023. PMID: 37393950. doi:10.1016/j.joen.2023.06.014
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。