金属伪影减少滤镜对完全与不完全牙根纵裂诊断准确性的影响
结果显示,e-vol DX BAR滤镜未改善两种骨折类型的AUC、准确率、敏感性和特异性(P > .05);不完全牙根纵裂的AUC、准确率和敏感性显著低于完全牙根纵裂(P < .0001)。
要点
- BAR滤镜未提升牙根纵裂的CBCT诊断准确性
- 根管内材料(牙胶尖、纤维桩、金属桩)未影响诊断
- 不完全牙根纵裂的敏感性显著低于完全牙根纵裂
核心结论
- BAR滤镜未改善牙根纵裂的诊断准确性,根管内材料也未影响诊断,不完全牙根纵裂与敏感性下降高度相关。
研究设计与样本
- 标本置于干燥人类颅骨内,在无充填、牙胶尖、纤维桩和金属桩4种情况下行Prexion 3D CBCT扫描。
- 用定制楔子在万能试验机上造成不完全牙根纵裂,复查确认后再完成骨折并再次扫描。
- 由e-Vol DX软件分别以BAR滤镜和原图评估牙根纵裂,用DeLong和McNemar检验比较AUC、准确率、敏感性和特异性。
主要发现
- BAR滤镜未改善两种骨折类型的AUC、准确率、敏感性和特异性(P > .05)。
- 不完全牙根纵裂的AUC、准确率和敏感性显著低于完全牙根纵裂(P < .0001)。
- 根管内材料未影响牙根纵裂的诊断。
临床怎么用
- 解读CBCT诊断牙根纵裂时不应依赖BAR滤镜提升准确性。
- 对疑似不完全牙根纵裂应警惕敏感性下降,必要时结合其他检查。
- 根管内存在牙胶尖、纤维桩或金属桩时,诊断准确性未见明显受其影响。
局限性
- 为体外研究,使用干燥人类颅骨与离体牙,与临床活体条件存在差异。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: This study aimed to assess the influence of the e-vol DX BAR filter on the diagnostic accuracy of complete and incomplete vertical root fractures (VRFs). METHODS: Twenty single-rooted teeth were selected. The tooth crowns were removed, and the root canals were prepared up to a 40/.06 instrument (ProTaper Next). Each specimen was scanned in a Prexion 3D cone-beam computed tomography (CBCT) device, in a dry human skull, in 4 different situations: no root canal filling, gutta-percha, fiberglass post, and metal post. The specimens were fractured in a universal testing machine by using a customized wedge. Each specimen was reinspected to confirm the presence of incomplete VRFs. Another set of CBCT scans was performed. Then, the fractures were completed in the universal testing machine, the teeth were reinspected, and the CBCT images were acquired again. Images were assessed by using the e-Vol DX software system, with the BAR filter and with the original images, for the diagnosis of VRFs. Statistical analysis was performed by using the DeLong and McNemar tests for the comparison of the area under the receiver operating characteristic curves (AUC), accuracy, sensitivity, and specificity values. RESULTS: The use of the BAR filter did not improve the diagnostic values of AUC, accuracy, sensitivity, and specificity in both fracture patterns (P > .05). Incomplete VRFs presented significantly lower AUC, accuracy, and sensitivity compared with complete VRFs (P < .0001). CONCLUSIONS: The BAR filter did not improve the diagnostic accuracy of VRFs. The intracanal materials also did not influence the diagnosis. Incomplete VRFs were highly associated with a decrease in sensitivity.
引用
- de Lima Dias-Junior LC, Correa M, Cesar Lyra Porto O, et al. Influence of a Metal Artifact Reduction Filter on the Diagnostic Accuracy of Complete and Incomplete Vertical Root Fractures. J Endod. 2023. PMID: 37479111. doi:10.1016/j.joen.2023.07.010
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。