文献速递口腔修复与种植

触觉操控与机器视觉控制协作机器人种植牙精度与安全性转化研究

Accuracy and safety of a haptic operated and machine vision controlled collaborative robot for dental implant placement: A translational study.
PubMed 文献 · Clin Oral Implants Res · 2023-06-12

该转化研究在树脂模型上测试锁扣结构后,将协作机器人辅助种植应用于21例患者(28枚种植体),单牙缺失手术中位时间23分钟,平台偏差均值0.54 mm、根尖偏差0.54 mm、角度偏差0.79°,下颌根尖偏差显著大于上颌,显示良好精度与安全性。

要点

  • 协作机器人辅助种植在模型与临床中均显示良好位置精度。
  • 单牙缺失中位手术时间23分钟,无牙颌为47和70分钟。
  • 下颌种植体根尖偏差显著大于上颌。

核心结论

  • 协作机器人辅助种植在体外模型和临床病例系列中均表现出优异的位置精度与安全性。
  • 仍需更多技术开发和临床研究以支持机器人手术在口腔种植中的引入。

研究设计与样本

  • 临床前模型研究在树脂牙弓模型上测试机械臂-手机锁扣结构。
  • 临床病例系列纳入21例患者共28枚种植体,其中2例无牙颌、19例单牙缺失。
  • 手术时间被记录,并测量种植体平台偏差、根尖偏差和角度偏差,分析影响因素。

主要发现

  • 体外模型使用锁扣结构时,平台偏差、根尖偏差和角度偏差均值(SD)分别为0.37(0.14)mm、0.44(0.17)mm和0.75(0.29)°。
  • 单牙缺失中位手术时间为23分钟(四分位距20-25分钟),两例无牙颌手术时间分别为47和70分钟。
  • 单牙缺失平台偏差、根尖偏差和角度偏差均值(SD)分别为0.54(0.17)mm、0.54(0.11)mm和0.79(0.22)°。
  • 无牙颌平台偏差、根尖偏差和角度偏差均值(SD)分别为0.53(0.17)mm、0.58(0.17)mm和0.77(0.26)°。
  • 下颌种植体的根尖偏差显著大于上颌。

临床怎么用

  • 协作机器人可用于单牙缺失及无牙颌的种植体植入,需关注下颌根尖偏差可能更大。
  • 临床应用中应结合模型验证的锁扣结构以保障精度。

局限性

  • 研究为临床前模型加小样本病例系列,仍需更多技术开发与临床研究支持机器人手术的引入。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Multiple generations of medical robots have revolutionized surgery. Their application to dental implants is still in its infancy. Co-operating robots (cobots) have great potential to improve the accuracy of implant placement, overcoming the limitations of static and dynamic navigation. This study reports the accuracy of robot-assisted dental implant placement in a preclinical model and further applies the robotic system in a clinical case series. MATERIALS AND METHODS: In model analyses, the use of a lock-on structure at robot arm-handpiece was tested in resin arch models. In a clinical case series, patients with single missing teeth or edentulous arch were included. Robot-assisted implant placement was performed. Surgery time was recorded. Implant platform deviation, apex deviation, and angular deviation were measured. Factors influencing implant accuracy were analyzed. RESULTS: The in vitro results showed that with a lock-on structure, the mean (SD) of platform deviation, apex deviation, and angular deviation were 0.37 (0.14) mm, 0.44 (0.17) mm, and 0.75 (0.29)°, respectively. Twenty-one patients (28 implants) were included in the clinical case series, 2 with arches and 19 with single missing teeth. The median surgery time for single missing teeth was 23 (IQ range 20-25) min. The surgery time for the two edentulous arches was 47 and 70 min. The mean (SD) of platform deviation, apex deviation, and angular deviation was 0.54 (0.17) mm, 0.54 (0.11) mm, and 0.79 (0.22)° for single missing teeth and for 0.53 (0.17) mm, 0.58 (0.17) mm, and 0.77 (0.26)° for an edentulous arch. Implants placed in the mandible had significantly larger apex deviation than those in the maxilla. CONCLUSION: Cobot-assisted dental implant placement showed excellent positional accuracy and safety in both the in vitro study and the clinical case series. More technological development and clinical research are needed to support the introduction of robotic surgery in oral implantology. Trial registered in ChiCTR2100050885.

来源:Clin Oral Implants Res,查看原文。版权归原作者所有。

引用

  1. Qiao SC, Wu XY, Shi JY, et al. Accuracy and safety of a haptic operated and machine vision controlled collaborative robot for dental implant placement: A translational study. Clin Oral Implants Res. 2023. PMID: 37309242. doi:10.1111/clr.14112

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。