文献速递口腔修复与种植

12种环境光照条件下共聚焦口内扫描仪准确度评估

Accuracy assessment (trueness and precision) of a confocal based intraoral scanner under twelve different ambient lighting conditions.
PubMed 文献 · J Dent · 2023-04-26

研究以实验室扫描仪获取的标准STL为参考,用共聚焦口内扫描仪PrimeScan在12种环境光照条件下各扫描10次,共120个STL文件,比较扫描畸变。结果显示1000 lux组扫描畸变最小(中位数±IQR 69.5±97.4 µm),0 lux组畸变最大(355.5±488.0 µm),环境光照条件影响该口内扫描仪的准确度。

要点

  • 1000 lux组扫描畸变最小,中位数±IQR为69.5±97.4 µm。
  • 0 lux组扫描畸变最大,为355.5±488.0 µm(p < 0.05)。
  • 环境光照条件影响共聚焦口内扫描仪PrimeScan的准确度。

核心结论

  • 环境光照条件影响所测试口内扫描仪的准确度;1000 lux时准确度最高,0 lux时扫描准确度最低。

研究设计与样本

  • 体外研究:用实验室扫描仪L2i扫描牙颌模型获得参考STL文件,由修复科医生用共聚焦口内扫描仪PrimeScan在12种环境光照条件下扫描,每组10个STL文件,共120个,用Meshlab软件比较畸变;因Shapiro-Wilk检验提示分布非正态,采用非参数Kruskal-Wallis及两两多重比较检验(α = 0.05)。

主要发现

  • 1000 lux组扫描畸变中位数±IQR最小,为69.5±97.4 µm。
  • 8000 lux组次之,中位数±IQR为166.5±318.1 µm。
  • 0 lx组畸变最高,为355.5±488.0 µm(p < 0.05)。

关键数值与推荐

  • 每组记录10个STL文件,共120个STL文件与参考STLr比较畸变。
  • 统计显著性水平α = 0.05,采用Kruskal-Wallis及两两多重比较检验。

临床怎么用

  • 原文未报告临床操作建议。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The ambient lighting condition has been identified as an important factor that influences the accuracy of intraoral scanners (IOSs). The purpose of this study was to evaluate the influence of 12 different ambient lighting conditions on the accuracy of a confocal based IOS (PrimeScan). MATERIALS AND METHODS: A typodont was digitized using a laboratory scanner (L2i) to obtain a reference standard tessellation language (STLr) file. A restorative dentist recorded the scans using an IOS (PrimeScan) under 12 different ambient lighting conditions where the luminosity was measured using a light meter (LX1330B Light Meter). Twelve groups were created, namely 0-, 500-, 1000-, 2000-, 3000-, 4000-, 5000-, 6000-, 7000-, 8000-, 9000-, and 10 000 lux groups. Ten STL files were recorded per group. The STLr file was used as a reference with which to compare the distortion of the 120 STL files obtained using a software program (Meshlab). The normality Shapiro-Wilk test indicated that the distributions were not normal. Therefore, the nonparametric Kruskal-Wallis and pairwise multicomparison tests were used to analyze the data (α = 0.05). RESULTS: The group with the 1000 lux lighting condition obtained the smallest median ±interquartile range (IQR) with scanning distortion values of 69.5 ± 97.4 µm, followed by the 8000 lux group with a median ±IQR of 166.5 ± 318.1 µm. The 0 lx group presented the highest distortion values with a mean ±IQR of 355.5 ± 488.0 µm (p < 0.05). CONCLUSIONS: Ambient lighting conditions influenced the accuracy of the IOS tested. The highest accuracy values were obtained with 1000 lux. The lowest scanning accuracy was obtained with 0 lux.

来源:J Dent,查看原文。版权归原作者所有。

引用

  1. Piedra-Cascón W, Adhikari RR, Özcan M, et al. Accuracy assessment (trueness and precision) of a confocal based intraoral scanner under twelve different ambient lighting conditions. J Dent. 2023. PMID: 37116740. doi:10.1016/j.jdent.2023.104530

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。