扫描角度对PTR/LUM检测非龋洞性龋损的影响

Impact of angle on photothermal radiometry and modulated luminescence (PTR/LUM) value.
PubMed 文献 · J Dent · 2023-04-02

研究评估扫描角度对光热辐射测量与调制发光(PTR/LUM,Canary System)检测/量化非龋洞性龋损的影响,并分析PTR/LUM值与病损深度(LD)及病损下健全组织厚度(ST)的关联。结果显示角度间PTR/LUM值有显著差异,但检测性能无显著差异;PTR/LUM值与最深病损深度及LD/ST比值呈中度相关。

要点

  • 扫描角度在20°以内变化可统计学影响PTR/LUM值,但不影响其检测性能。
  • PTR/LUM值与非龋洞性病损深度呈正相关,且不受病损下健全组织厚度影响。
  • 临床纵向扫描难以保持同一位点同一角度,标准化这些参数很重要。

核心结论

  • 扫描角度在20°增量内可能统计学影响PTR/LUM值,但不影响PTR/LUM检测性能。
  • PTR/LUM值与非龋洞性病损深度呈正相关,且不受病损下健全组织厚度影响。

研究设计与样本

  • 每颗牙分别以90°直接接触非龋洞性病损或健全光滑面中心,及向颊/舌/𬌗/颈四个方向倾斜10°和20°扫描,48小时后重复。
  • 在μ-CT图像同一扫描方向测量病损深度(LD)和ST(ST=5000 µm最大PTR/LUM扫描深度-LD)。

主要发现

  • PTR/LUM值在不同扫描角度间显示显著差异。
  • 总体敏感度(78%-89%)、特异度(66%-87%)、AUC(0.86-0.92)及ICC(0.89-0.99)在不同角度间无显著差异;基于病损扩展范围的敏感度在不同角度间亦无显著差异。
  • PTR/LUM值与最深病损深度及LD/ST比值呈中度相关(0.56-0.74)。

临床怎么用

  • 临床纵向测量/扫描同一位点同一角度具有挑战性,标准化这些参数很重要。
  • 偏离垂直方向20°以内扫描不影响PTR/LUM检测性能,且PTR/LUM值与龋坏深度呈正中度相关。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To evaluate the impact of scanning angles to detect/quantify non-cavitated caries by photothermal-radiometry and modulated-luminescence (PTR/LUM, Canary System) and to evaluate the association of PTR/LUM value with lesion depth (LD), including sound tissue thickness under the lesion (ST). METHODS: Thirty human extracted premolars were selected based on micro-computed tomography [μ-CT: sound (n=12), lesions into outer-half of enamel (n=6), lesions into inner-half of enamel (n=6), lesions into outer one-third of dentine (n=6)]. Each tooth sample was scanned 90° directly contacted to the center of non-cavitated lesion or sound smooth surface, and tilted 10° and 20° in four directions: buccal/lingual/occlusal/cervical. The procedure was repeated 48 h later. Lesion depth and ST [ST=5000 µm (maximum PTR/LUM scanning depth)-LD] were measured at the same scanning direction on μ-CT images. Sensitivity, specificity, area under the Receiver Operating Characteristic curve (AUC), and intraclass correlation coefficients (ICC) for different scanning angles were calculated. Sensitivity was further evaluated based on lesion extensions. Relationships between PTR/LUM value and lesion depth, and between PTR/LUM value and LD/ST-Ratio were evaluated. RESULTS: PTR/LUM value showed significant differences among scanning angles. Overall sensitivity (78%-89%), specificity (66%-87%), AUC (0.86-0.92) and ICC (0.89-0.99), sensitivity based on lesion extensions presented no significant differences among angles. PTR/LUM value showed moderate correlations (0.56-0.74) with deepest lesion depth and LD/ST-Ratios. CONCLUSION: The scanning angle within 20° increments might impact PTR/LUM value statistically; however, it did not affect PTR/LUM detection performance. PTR/LUM values were positively correlated with non-cavitated lesion depth, and not affected by sound tissue thickness under the lesion. CLINICAL SIGNIFICANCE: Clinically, it is challenging to measure/scan at the same location and same angle longitudinally, however, it is important to standardize these parameters. Scanning within 20° deviation from perpendicular did not affect detection performance of PTR/LUM, and PTR/LUM value showed positive moderate correlation with caries depth.