无套管导板牙髓病治疗精度体外研究及复杂上颌侧切牙病例
本研究在3D打印颌骨模型上评估无套管导板用于严重根管钙化(PCO)根管治疗开髓的精度,由两名不同经验水平的操作者共完成88个导板开髓洞(每人44个)。平均开髓洞长度15.3 mm,冠部与根尖偏移分别为0.5 mm和0.7 mm,平均角度偏差1.5°,三名测量参数在操作者间均无统计学显著差异。文中另报告一例复杂上颌侧切牙的临床治疗以展示无套管导板的应用。
要点
- 无套管导板在3D打印颌骨模型上开髓,平均冠部偏移0.5 mm、根尖偏移0.7 mm、角度偏差1.5°。
- 两名不同经验操作者各完成44个开髓洞,三项精度参数均无统计学显著差异。
- 作者认为在其研究局限内,无套管导板是引导牙髓治疗的一种精确方法。
核心结论
- 本研究在其局限内表明,无套管导板是引导牙髓治疗的一种精确方法。
- 使用该导板时,不同操作者之间未发现统计学显著差异。
研究设计与样本
- 体外研究:选取上、下颌各一个锥形束CT(CBCT)容积,设计并3D打印严重根管钙化(PCO)模型。
- 对右侧至左侧第二前磨牙进行开髓洞虚拟规划,模型安装于仿头模以模拟真实患者。
- 两名牙髓病经验水平不同的操作者完成导板下开髓,共在8个3D打印模型上钻取88个导板开髓洞(每名操作者44个)。
- 术后拍摄CBCT进行分析。
主要发现
- 共在8个3D打印模型上完成88个导板开髓洞,每名操作者44个。
- 开髓洞平均长度为15.3 mm。
- 平均冠部偏移为0.5 mm,平均根尖偏移为0.7 mm。
- 平均角度偏差为1.5°。
- 三名测量参数在操作者之间均未发现统计学显著差异。
临床怎么用
- 作者认为,该方法为传统牙髓导板提供了一种有价值的替代选择,精度结果相似。
局限性
- 原文在结论中写明「within its limitations」,未具体列出各项局限内容。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: The present study aims to assess the accuracy of sleeveless guided endodontics for root canal treatment of severe pulp canal obliteration (PCO) in 3D printed jaws. Additionally, the treatment of a complex lateral incisor is presented to illustrate the use of sleeveless guides in a clinical situation. METHODS: Two cone-beam computed tomography (CBCT) volumes of an upper and lower jaw were selected to design 3D printed models with PCO. Virtual planning of the access cavities was performed from right to left second premolar. Then, the models were mounted into a phantom head to simulate an actual patient. Two operators with different levels of experience in endodontics performed guided access cavities. The handpiece was guided by guiding rails placed against each other on the sides of the tooth. A post-operative CBCT scan was taken for analysis. RESULTS: Eighty-eight guided access cavities (44 per operator) were drilled on eight 3D printed models. The mean length of the access cavities was 15.3 mm, with a mean coronal and apical deviation of 0.5 mm and 0.7 mm respectively. The mean angular deviation was 1.5°. No statistically significant difference was found between operators for the three measured parameters. CONCLUSIONS: This study demonstrates, within its limitations, that sleveless guides represent an accurate method for guided endodontic treatment. No statistically significant difference between operators was found when using the guide. CLINICAL SIGNIFICANCE: This method offers a valuable alternative to conventional endodontic guides with similar accuracy results.