文献速递口腔修复与种植牙体牙髓病

不同牙体预备完成方式及即刻牙本质封闭对口腔扫描仪扫描精度的影响

The effect of different tooth preparation finishing procedures and immediate dentin sealing on the scanning accuracy of different intraoral scanners.
PubMed 文献 · J Dent · 2023-01-20

本体外研究在离体下颌磨牙上比较超粗砂、细砂、氧化铝喷砂和即刻牙本质封闭(IDS)四种牙体预备完成方式对4种口内扫描仪(IOS)扫描精度的影响。结果显示完成方式和扫描仪类型均显著影响精度(P<.001),喷砂组精度最高,IDS组最低;i700组精度最高,CS与Trios组最低。作者认为喷砂可能被推荐用于最大化扫描精度。

要点

  • 牙体预备完成方式和IOS类型均显著影响扫描真度与精度(P<.001)。
  • 喷砂组真度与精度最高,IDS组最低;i700组最高,CS与Trios组最低。
  • 作者认为喷砂可能被推荐用于最大化所测IOS的扫描精度。

核心结论

  • 不同牙体预备完成方式影响4种口内扫描仪的扫描精度,喷砂方式获得最佳精度值。
  • 所有亚组间真度差异为19µm,精度差异为4.69µm。

研究设计与样本

  • 体外研究:在离体下颌磨牙上以超粗金刚砂车针完成全冠牙体预备。
  • 按完成方式分4组:超粗砂(150µm车针,SCG组)、细砂(30µm车针,FG组)、27µm氧化铝喷砂(APA组)、即刻牙本质封闭(IDS组)。
  • 每组按扫描系统分5亚组:实验室扫描仪(对照亚组,T710;Medit)、Trios 4(Trios亚组)、CS 3800(CS亚组)、i700 wireless(i700亚组)、iTero Element 5D Plus(iTero亚组),n=20。
  • 以最佳拟合算法和工程软件(Geomagic Control X)将对照文件与各实验扫描对齐,计算均方根(RMS)误差。
  • 采用双因素方差分析和Tukey检验(α=.05)。

主要发现

  • 牙体预备完成方式(P<.001)和所评估的IOS(P<.001)是真度与精度的显著预测因素。
  • APA组真度与精度最高,IDS组真度与精度最低。
  • i700亚组真度与精度最高,CS与Trios亚组真度与精度最低。

关键数值与推荐

  • 超粗砂组使用砂粒尺寸150µm的车针,细砂组使用砂粒尺寸30µm的车针,喷砂组使用27µm氧化铝颗粒。
  • 每个亚组样本量为n=20。
  • 所有亚组间真度差异为19µm,精度差异为4.69µm。

临床怎么用

  • 作者认为,所用牙体预备完成方式可降低所测任一口内扫描仪的扫描精度。
  • 作者认为,喷砂完成方式可能被推荐用于最大化所测IOS的扫描精度。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To measure the effect of different tooth preparation finishing procedures (super-coarse grit, fine grit, and air-particle abrasion) and immediate dentin sealing (IDS) on the scanning accuracy of 4 intraoral scanners (IOSs). MATERIAL AND METHODS: A tooth preparation for a full-coverage restoration was performed on an extracted mandibular molar using super-coarse diamond burs. Four groups were created depending on the tooth preparation finishing procedure: super-coarse grit (bur with a grit size of 150 µm) (SCG group), fine grit (bur with a grit size of 30 µm) (FG), air-particle abrasion with 27-µm aluminum oxide particles (APA group), and IDS (IDS group). Each group was divided into 5 subgroups according to the scanning system used to digitize the tooth preparation: laboratory scanner (control subgroup) (T710; Medit), Trios 4 (Trios subgroup), CS 3800 (CS subgroup), i700 wireless (i700 subgroup), and iTero Element 5D Plus (iTero subgroup) (n=20). For each subgroup, the control file was aligned with each experimental scan using the best-fit algorithm and an engineering program (Geomagic Control X). The discrepancy between the control and experimental files of each subgroup was computed by measuring the root mean square (RMS) error. Two-way ANOVA and Tukey tests were used to analyze the data (α=.05). RESULTS: Tooth preparation finishing procedures (P<.001) and the IOS assessed (P<.001) were significant predictors of the trueness and precision values obtained. The highest trueness and precision values were measured in the APA group, while the IDS group had the lowest trueness and precision. Additionally, the i700 subgroup obtained the highest trueness and precision values, while the CS and Trios subgroups had the lowest trueness and precision values. CONCLUSIONS: The different tooth preparations finishing procedures tested influenced on the scanning accuracy of the 4 IOSs considered. The air-particle abrasion procedure obtained the best accuracy values. The trueness discrepancies measured among all the subgroups was 19µm and the precision discrepancies measured among all the subgroups was 4.69 µm. CLINICAL SIGNIFICANCE: The tooth preparation finishing procedure used can reduce the intraoral scanning accuracy of any of the intraoral scanners tested. The air-particle abrasion finishing procedure might be recommended for maximizing the scanning accuracy of the IOSs tested.

来源:J Dent,查看原文。版权归原作者所有。

引用

  1. Revilla-León M, Cascos-Sánchez R, Barmak AB, et al. The effect of different tooth preparation finishing procedures and immediate dentin sealing on the scanning accuracy of different intraoral scanners. J Dent. 2023. PMID: 36682722. doi:10.1016/j.jdent.2023.104431

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。