无线电子根尖定位仪活体准确性的微型CT研究

Micro-CT Study of the In Vivo Accuracy of a Wireless Electronic Apex Locator.
PubMed 文献 · J Endod · 2022-06-30

研究纳入5例患者计划拔除的11颗活髓牙,在根管预备后用Root ZX II与Wirele-X两种电子根尖定位仪(EAL)测定工作长度,并以微型CT双扫描配准计算文件尖端与主根尖孔的距离误差。在±0.5 mm容差内,两种定位仪的绝对距离值无显著差异(P=.82),检出主根尖孔位置的能力亦无显著差异(χ2=0.2588;P=.6109);Root ZX II与Wirele-X的准确性分别为81.8%与90.9%。结论认为两者在活体检测主根尖孔方面表现相似。

要点

  • 在±0.5 mm容差内,两种EAL的绝对距离值无显著差异(P=.82)。
  • 检出主根尖孔位置的能力无显著差异(χ2=0.2588;P=.6109)。
  • Root ZX II与Wirele-X的准确性分别为81.8%与90.9%。

核心结论

  • Root ZX II与Wirele-X在活体检测主根尖孔方面表现相似。
  • 采用严格标准时,Root ZX II与Wirele-X的准确性分别为81.8%与90.9%。

研究设计与样本

  • 研究纳入5例患者计划拔除的11颗活髓牙,比较Wirele-X与Root ZX II两种电子根尖定位仪在活体测定主根尖孔位置的准确性。
  • 根管预备后使用08与10号K锉疏通至根尖三分之一,并以2.5%次氯酸钠冲洗。
  • 用K型锉配合Root ZX II与Wirele-X测定工作长度,直至数值显示为「0.0」,固定硅胶止动片后拔牙。
  • 拔除牙以双扫描方案行微型CT成像,将含锉与不含锉的图像序列配准,计算锉尖与主根尖孔之间的绝对差值作为测量误差。
  • 锉尖超出主根尖孔记为正值、未达主根尖孔记为负值;准确性判定为锉尖未超出主根尖孔且测量稳定在±0.5 mm以内。

主要发现

  • 在±0.5 mm容差内,两种受试EAL的绝对距离值无显著差异(P=.82)。
  • 在±0.5 mm容差内,两种EAL检出主根尖孔位置的能力无显著差异(χ2=0.2588;P=.6109)。
  • Root ZX II在±0.5 mm内的准确性为81.8%,Wirele-X为90.9%。

关键数值与推荐

  • 准确性判定标准为锉尖未超出主根尖孔且测量稳定在±0.5 mm以内。
  • 锉尖超出主根尖孔记为正值,未达主根尖孔记为负值。
  • 测量误差为锉尖与主根尖孔之间的绝对差值。
  • 显著性水平设为5%。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: This study aimed to compare the in vivo accuracy of Wirele-X and RootZX II electronic apex locators (EALs) in determining the position of the major foramen using micro-computed tomography (micro-CT) as the analytical tool. METHODS: Eleven vital teeth planned for extraction from 5 patients were used. After conventional access cavity preparation, root canals were flared and negotiated up to the apical third with sizes 08 and 10 K-files followed by irrigation with 2.5% NaOCl. K-type files were used to determine the working length of the selected canals using Root ZX II and Wirele-X apex locators until their numeric displays read "0.0." After fixing the silicon stop to the file, teeth were extracted and imaged in a micro-CT device using a double-scan protocol. Image stacks, with and without the file in the root canal, were then co-registered and the measurement error calculated as the absolute difference between the tip of the file and the major foramen. Positive and negative values were recorded when the file tip was detected beyond or short of the major foramen, respectively. Accuracy was determined on stable measurements within ± 0.5 mm when the file tip did not extend beyond the major foramen. The χ2 test was applied to compare the ability of the EALs to detect the position of the major foramen, and the t test for dependent variables was used to verify differences in the 2 measurements obtained in each tooth. Significance level was set at 5%. RESULTS: Within a tolerance level of ± 0.5 mm, no significant differences were observed between the tested EALs regarding the absolute distance values (P = .82) or in their ability to detect the position of the major foramen (χ2 = 0.2588; P = .6109). The accuracy of the Root ZX II and the Wirele-X apex locators within ± 0.5 mm were 81.8% and 90.9%, respectively. CONCLUSIONS: Root ZX II and Wirele-X performed similarly regarding the in vivo detection of the major foramen. Using strict criteria, the accuracy of the Root ZX II and the Wirele-X apex locators were 81.8% and 90.9%, respectively.