文献速递牙体牙髓病

TruNatomy与ProTaper Gold器械对下颌磨牙根尖周牙本质保存及根尖区管腔扩大的影响

The Impact of TruNatomy and ProTaper Gold Instruments on the Preservation of the Periradicular Dentin and on the Enlargement of the Apical Canal of Mandibular Molars.
PubMed 文献 · J Endod · 2022-02-15

本研究将20颗下颌磨牙经显微CT扫描后解剖配对分为两组(每组10颗),分别使用TruNatomy和ProTaper Gold器械预备根管,比较根尖周牙本质保存与根尖区管腔扩大情况。结果显示两组在未预备根管壁面积和牙本质厚度减少上无差异(P>.05),所有组根管偏移均小于0.1 mm,仅在近中颊根管根尖三分之一处TruNatomy组偏移值更低,而ProTaper Gold在近中根冠部去除牙本质更多(分别为1.8%和1.0%,P<.05)。结论为两种系统均能有效预备下颌磨牙根管,在未触及根管壁和剩余牙本质厚度方面相似,仅在近中根管根尖偏移和冠三分之一牙本质去除百分比上略有差异,但无临床显著错误。

要点

  • 两组在未预备根管壁面积和牙本质厚度减少上无显著差异(P>.05)。
  • 所有组根管偏移均小于0.1 mm,仅近中颊根管根尖三分之一处TruNatomy组偏移值更低。
  • ProTaper Gold在近中根冠部去除牙本质多于TruNatomy(1.8%和1.0%,P<.05)。

核心结论

  • TruNatomy和ProTaper Gold均能有效预备下颌磨牙根管,在未触及根管壁和剩余牙本质厚度方面相似,仅在近中根管根尖偏移和冠三分之一牙本质去除百分比上略有差异,但无临床显著错误。

研究设计与样本

  • 将20颗下颌磨牙经显微CT扫描后解剖配对,分为2组(n=10),分别使用TruNatomy和ProTaper Gold器械预备根管。

主要发现

  • 两组在未预备根管面积和牙本质厚度减少上无差异(P>.05)。
  • 所有组根管偏移均小于0.1 mm,仅在近中颊根管根尖三分之一处TruNatomy组偏移值更低。
  • ProTaper Gold在近中根冠部去除牙本质多于TruNatomy(分别为1.8%和1.0%,P<.05)。

关键数值与推荐

  • ProTaper Gold组近中根管预备至F2(25/.08v),远中根管预备至F3(30/.09v)。
  • TruNatomy组近中根管扩大至prime(26/.04v),远中根管扩大至medium(36/.03v)。
  • 统计学检验采用Mann-Whitney检验、Student t检验和非度量多维尺度分析,α设为5%。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: This study aimed to evaluate the preservation of periradicular dentin and the enlargement of the apical canal of mandibular molars with TruNatomy (Dentsply Sirona, Ballaigues, Switzerland) and ProTaper Gold (Dentsply Sirona) instruments. METHODS: Twenty mandibular molars were scanned in a micro-computed tomographic device, anatomically paired, and distributed into 2 groups (n = 10). In the ProTaper Gold group, mesial and distal canals were prepared up to F2 (25/.08v) and F3 (30/.09v) instruments, whereas in the TruNatomy group, mesial and distal canals were enlarged up to the prime (26/.04v) and medium (36/.03v) instruments, respectively. After a new scan, the surface area, volume, unprepared areas, transportation, percentage of dentin removal, and dentin thickness parameters were calculated. Data were compared between groups using the Mann-Whitney test, the Student t test, and the nonmetric multidimensional scaling test with alpha set at 5%. RESULTS: No difference was found between groups regarding unprepared canal areas and the reduction of dentin thickness (P > .05). Transportation was lower than 0.1 mm in all groups, and statistical differences were observed only at the apical third of the mesiobuccal canal with lower values in the TruNatomy group. ProTaper Gold removed more dentin than TruNatomy at the coronal level of mesial roots (1.8% and 1.0%, respectively) (P < .05). CONCLUSIONS: TruNatomy and ProTaper Gold were efficient for performing canal preparation in mandibular molars. The tested systems were similar in terms of untouched canal walls and remaining dentin thickness and slightly different in the apical transportation of mesial canals and the percentage of dentin removal at the coronal third but without clinically significant errors.

来源:J Endod,查看原文。版权归原作者所有。

引用

  1. Silva EJNL, Lima CO, Barbosa AFA, et al. The Impact of TruNatomy and ProTaper Gold Instruments on the Preservation of the Periradicular Dentin and on the Enlargement of the Apical Canal of Mandibular Molars. J Endod. 2022. PMID: 35181453. doi:10.1016/j.joen.2022.02.003

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。