基于通用坐标系统比较虚拟技术登记单颗种植体位置:一项体外研究
本体外研究以30个含上颌前牙区种植体的树脂模型为对象,用立方角定义的通用坐标系统,将台式扫描仪、锥形束CT、口内扫描仪及石膏模型扫描与坐标测量机物理金标准比较。结果显示口内扫描仪在种植体颈部位置的登记精度优于石膏模型扫描。
要点
- 以坐标测量机物理测量为金标准,比较四种虚拟技术登记单颗种植体位置。
- 口内扫描仪在种植体颈部Δz和r上显著小于石膏模型扫描(p=0.006)。
- 各技术在种植体根尖位置的坐标和距离上无显著差异。
核心结论
- 物理测量可作为评估单颗种植体位置的金标准。
- 口内扫描仪在登记单颗种植体颈部位置方面比石膏模型扫描更准确。
研究设计与样本
- 体外研究,制备30个上颌无牙区树脂模型,于切牙区植入种植体。
- 每个模型制作带1 cm×1 cm×1 cm立方角的牙支持式丙烯酸支架。
- 以坐标测量机参照立方角物理测量种植体颈部和根尖的x、y、z坐标作为金标准。
- 树脂模型分别用台式扫描仪(BS组)、锥形束CT(CBCT组)和口内扫描仪(IOS组)扫描。
- 用开窗式聚醚印模灌注的石膏模型经台式扫描仪扫描(BS-cast组)。
主要发现
- BS组种植体颈部和根尖位置与坐标测量机结果存在统计学差异,r>0(p<0.001)。
- IOS组种植体颈部的Δz和r显著小于BS-cast组(p=0.006)。
- BS组、CBCT组、IOS组和BS-cast组在种植体根尖的坐标和距离上无显著差异。
临床怎么用
- 原文未据此给出临床推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: The aim of this in-vitro study was to compare the virtual techniques for registering single-implant position to the physical gold standard using a universal-coordinate system. MATERIALS AND METHODS: Thirty dentate maxillary resin models with a dental implant inserted in the incisor region were prepared. On each model, a tooth-supported acrylic stent with a 1 cm x 1 cm x 1 cm cubic-corner (CC) was prepared. The Cartesian x,y,z-coordinate of the implant neck and apex were measured physically by a coordinate-measuring machine (CMM) with reference to this CC and served as the gold-standard. The resin models were scanned by a benchtop scanner (Group BS), cone-beam computed tomography (Group CBCT), and intraoral scanner (Group IOS). Stone casts, poured from open-tray polyether impression of the resin models, were scanned by the benchtop scanner (Group BS-cast). The implant neck and apex coordinates with reference to the CC were measured and the differences in the coordinates (∆x, ∆y, ∆z) and distance r from the gold standard were calculated. The data were analyzed by one-sample t-test and one-way ANOVA/Kruskal-Wallis test with the level of significance set at 0.05. RESULTS: The implant neck and apex positions of Group BS were statistically different from that of the CMM, r>0 (p<0.001). Group IOS showed a significant less ∆z and r at the implant neck than Group BS-cast (p = 0.006). No significant difference was found in the coordinates and distance at implant apex among Groups BS, CBCT, IOS and BS-cast. CONCLUSIONS: The physical measurements could be adopted as the gold standard in assessing the single-implant positions. The IOS was more accurate in registering the single-implant neck positions than scanning of the cast. CLINICAL SIGNIFICANCE: A universal-coordinate system defined by the cubic-corner allows comparing the virtual techniques in registering single-implant positions to the physical gold standard.
引用
- Pan Y, Heng C, Wu ZJ, et al. Comparison of the virtual techniques in registering single implant position with a universal-coordinate system: An in vitro study. J Dent. 2021. PMID: 34929339. doi:10.1016/j.jdent.2021.103925
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。