文献速递口腔修复与种植

评估视野内外金属物体产生的锥形束CT伪影

Evaluation of Cone-beam Computed Tomography Artifacts Produced by Metal Objects Located Within and Outside the Field of View.
PubMed 文献 · J Endod · 2021-12-08

本研究在人类尸体下颌骨一侧放置4个钛种植体尺寸的杆和4个氧化锆冠尺寸的圆盘,使用Accuitomo 170 CBCT机进行小视野和大视野扫描各9次,比较视野外金属物体在小视野图像中产生的伪影与大视野图像中金属物体位于视野内时产生的伪影。结果显示,随着金属物体数量增加,小视野和大视野切片的标准差均增加;小视野和大视野的标准差均值分别为3.6和2.5(P = .0000);53%的小视野切片在主观分析中伪影更多;100%的小视野切片被评定为更高质量。结论认为,对侧象限视野外的金属物体会影响小视野图像质量,但小视野图像分辨率高于大视野图像。

要点

  • 对侧象限视野外金属物体会影响小视野CBCT图像质量。
  • 小视野与大视野切片标准差均值分别为3.6与2.5(P = .0000)。
  • 100%的小视野切片被评定为更高质量。

核心结论

  • 对侧象限视野外的金属物体会影响小视野图像质量,但小视野图像分辨率高于大视野图像。

研究设计与样本

  • 在人类尸体下颌骨一侧放置4个钛种植体尺寸的杆和4个氧化锆冠尺寸的圆盘,使用Accuitomo 170 CBCT机共进行18次扫描(小视野9次、大视野9次),每次扫描取10个轴位切片转入ImageJ软件分析。
  • 由两名盲法观察者主观评定图像诊断质量和伪影存在情况。

主要发现

  • Wilcoxon符号秩检验显示,随着金属物体数量增加,小视野和大视野切片的标准差均增加。
  • 小视野和大视野的标准差均值分别为3.6和2.5(P = .0000)。
  • 主观分析中53%的小视野切片伪影更多。
  • 100%的小视野切片被评定为更高质量。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: Artifacts created by the presence of metal objects in the jaw negatively affect the cone-beam computed tomographic image quality. This study compares artifacts produced by metal objects outside of the field of view (FOV) in a small FOV CBCT image with those produced in a large FOV image in which the metal object is within the FOV. METHODS: We methodically placed 4 titanium implant-sized rods and 4 zirconium crown-sized disks on 1 side of a human cadaver mandible. Using the Accuitomo 170 CBCT machine (J Morita, Irvine, CA), a total of 18 scans (9 with a small FOV and 9 with a large FOV) were made. Ten axial slices from each scan were transferred to ImageJ software (National Institutes of Health, Bethesda, MD) for analysis. The mean standard deviation of all voxel values of a fixed region of interest (ie, uniform air located lingual to tooth #30) was compared between small and large FOV slices. Two blinded observers subjectively rated the images for diagnostic quality and the presence of artifacts. RESULTS: The Wilcoxon signed rank test showed that the standard deviation for both small and large FOV slices increases as the number of metal objects increases. The mean of the standard deviation for small and large FOVs is 3.6 and 2.5, respectively (P = .0000). Fifty-three percent of the small FOV slices had more artifacts in the subjective analysis. One hundred percent of the small FOV slices were rated as higher quality. CONCLUSION: Metal objects outside of the FOV in the contralateral quadrant do affect the quality of small FOV images. However, small FOV images have a higher resolution compared with large FOV images.

来源:J Endod,查看原文。版权归原作者所有。