HVAD 更换为 HeartMate 3 的装置交换:STS Intermacs 分析
基于 STS Intermacs 数据库,比较 HVAD 原发植入、HVAD 更换为 HeartMate 3 及 HVAD 更换为 HVAD 三组的生存。HVAD 更换为 HeartMate 3 较继续 HVAD 支持生存显著降低;但在需泵更换患者中,更换为 HeartMate 3 生存优于更换为 HVAD。
要点
- HVAD 更换为 HeartMate 3 较继续 HVAD 支持生存显著降低
- 需泵更换者中,换为 HeartMate 3 生存优于换为 HVAD
- 目前证据不足以支持择期将 HVAD 更换为 HeartMate 3
核心结论
- 与继续 HVAD 支持相比,更换为 HeartMate 3 与死亡率显著增加相关。
- 对于需要泵更换的 HVAD 支持患者,更换为 HeartMate 3 显示更优生存。
- 目前证据不足以支持择期将 HVAD 更换为 HeartMate 3。
研究设计与样本
- 基于 STS Intermacs 数据库的三队列分析:原发 HVAD 植入队列(2017 年 1 月至 2021 年 3 月,n=3797)。
- HVAD 更换为 HVAD 队列(2017 年 1 月至 2021 年 3 月,n=234)。
主要发现
- HVAD 更换为 HeartMate 3 后 6 个月生存率为 73.8%(70%CI 68.6-77.8),继续 HVAD 支持为 79.0%(70%CI 78.3-79)。
- HVAD 更换为 HeartMate 3 后 1 年生存率为 85.9%(70%CI 79.5-90.5),HVAD 更换为 HVAD 为 66.6%(70%CI 63.0-70.0),P=.009。
临床怎么用
- 对 HVAD 支持患者,不应择期更换为 HeartMate 3。
- 当 HVAD 患者需泵更换时,更换为 HeartMate 3 可能优于更换为 HVAD。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: On June 3, 2021 Medtronic, Inc announced discontinuation of the HVAD left ventricular assist device. The purpose of this analysis was to provide summary data on surgical risks of HVAD to HeartMate 3 exchange and compare survival after HVAD to HeartMate 3 exchange to survival after primary HVAD implantation. METHODS: Three cohorts within The Society of Thoracic Surgeons Intermacs database were identified: primary HVAD implant cohort (January 2017 to March 2021, n = 3797), HVAD to HeartMate 3 exchange cohort (December 2017 to March 2021, n = 45), and HVAD to HVAD exchange cohort (January 2017 to March 2021, n = 234). Mortality after HVAD to HeartMate 3 exchange was modeled and compared with the constant hazard phase for risk of mortality while on continued HVAD support. As a secondary analysis outcomes and survival were compared between patients who underwent HVAD to HeartMate 3 and HVAD to HVAD exchange. RESULTS: HVAD to HeartMate 3 exchange was associated with significantly reduced survival compared with survival while remaining on HVAD support (6 months after exchange, 73.8% [70% confidence interval, 68.6-77.8] vs 79.0% [70% confidence interval, 78.3-79] for continued HVAD support). Compared with HVAD to HVAD exchange, survival was higher after replacement with HeartMate 3 (1 year: 85.9% [70% confidence interval, 79.5-90.5] vs 66.6% [70% confidence interval, 63.0-70.0], P = .009). CONCLUSIONS: Compared with continued support on HVAD, an exchange to HeartMate 3 was found to be associated with a significant increase in mortality. For patients who required pump exchange on HVAD support, exchange to HeartMate 3 demonstrated superior survival. Currently there is insufficient evidence to support elective exchange from an HVAD to HeartMate 3.
引用
- Cogswell R, Cantor RS, Vorovich E, et al. HVAD to Heartmate 3 Device Exchange: A Society of Thoracic Surgeons Intermacs Analysis. Ann Thorac Surg. 2021. PMID: 34678286. doi:10.1016/j.athoracsur.2021.09.031
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。