文献速递瓣膜外科

二尖瓣与三尖瓣成形环裂开的系统综述与汇总分析

Mitral and tricuspid annuloplasty ring dehiscence: a systematic review with pooled analysis.
PubMed 文献 · Eur J Cardiothorac Surg · 2021-01-01

二尖瓣裂开率因环类型显著不同(半硬质1.86%、硬质2.32%、柔性0.43%,P<0.001),缺血性二尖瓣反流裂开率高于非缺血性(3.91% vs 1.63%,P=0.022)。

要点

  • 二尖瓣成形环总体裂开率1.43%,中位诊断时间术后4.5±1.0个月。
  • 硬质环二尖瓣裂开率2.32%高于半硬质1.86%和柔性0.43%(P<0.001)。
  • 缺血性二尖瓣反流裂开率3.91%高于非缺血性1.63%(P=0.022)。

核心结论

  • 硬质、半硬质和柔性成形环均提供可接受的瓣膜修复结果,但二尖瓣成形环裂开在硬质环中临床更常见。
  • 理解环裂开的多因素性质有助于识别高危患者并改善临床结局。

主要发现

  • 二尖瓣裂开87例、三尖瓣裂开30例,总体裂开率1.43%,中位诊断时间4.5±1.0个月。
  • 二尖瓣裂开率因环类型显著不同:半硬质1.86%、硬质2.32%、柔性0.43%(P<0.001)。
  • 二尖瓣裂开率因环尺寸(P=0.067)和形状(P=0.281)无显著差异。
  • 缺血性二尖瓣反流裂开率3.91%高于非缺血性1.63%(P=0.022)。
  • 三尖瓣研究中10项有9项未报告任何裂开。

关键数值与推荐

  • 二尖瓣成形环裂开率:半硬质1.86%、硬质2.32%、柔性0.43%(P<0.001)。
  • 缺血性 vs 非缺血性二尖瓣反流裂开率:3.91% vs 1.63%(P=0.022)。
  • 二尖瓣环尺寸与裂开率无关(P=0.067),环形状与裂开率无关(P=0.281)。

临床怎么用

  • 选择二尖瓣成形环时需权衡环类型相关的裂开风险,硬质环裂开率较高。
  • 对缺血性二尖瓣反流患者应加强术后随访,警惕成形环裂开。
  • 术后早期(约4.5个月)是成形环裂开诊断的中位时间,需在此阶段关注复发反流。

局限性

  • 纳入研究间可能存在异质性,汇总分析结果需谨慎解读。
  • 三尖瓣研究多数未报告裂开,三尖瓣裂开率数据有限。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Mitral and tricuspid ring annuloplasty dehiscence with consequent recurrent valve regurgitation is a rare but challenging procedural failure. The incidence and predisposing risk factors for annuloplasty ring dehiscence include technical and pathological ones. METHODS: A systematic database search with pooled analysis was conducted of original articles that only included dehiscence rate of mitral and tricuspid ring in EMBASE, MEDLINE, Cochrane database and Google Scholar, from inception to November 2020. The outcomes included were dehiscence rate in mitral and tricuspid, type of ring implanted, dehiscence rate by pathology and by ring size and shape. RESULTS: Our search yielded 821 relevant studies. Thirty-three studies met the inclusion criteria with a total of 10 340 patients (6543 mitral, 1414 tricuspid) of which 87 (mitral) and 30 (tricuspid) had dehiscence. Overall, dehiscence rate was 1.43%, diagnosed at a median of 4.5 ± 1.0 months postoperatively. A significant difference in mitral dehiscence rate was found by ring type (semi-rigid 1.86%, rigid 2.32%; flexible 0.43%; P < 0.001). There was no significant difference in rate of dehiscence by ring size (P = 0.067) and shape in mitral (P = 0.281) but there was higher dehiscence rate in ischaemic compared to non-ischaemic mitral regurgitation (3.91% vs 1.63%; P = 0.022). Among tricuspid studies, 9 of 10 studies did not report any dehiscence. CONCLUSIONS: Although rigid, semi-rigid and flexible annuloplasty rings provide acceptable valve repair outcomes, mitral annuloplasty ring dehiscence is clinically more common among rigid rings. Understanding the multifactorial nature of ring dehiscence will help in identifying the patients at high risk and improve their clinical outcomes.

来源:Eur J Cardiothorac Surg,查看原文。版权归原作者所有。

引用

  1. Arjomandi Rad A, Naruka V, Vardanyan R, et al. Mitral and tricuspid annuloplasty ring dehiscence: a systematic review with pooled analysis. Eur J Cardiothorac Surg. 2021. PMID: 33880496. doi:10.1093/ejcts/ezab178

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。