HeartMate 3泵位置与流入角连续评估及其对不良事件的影响
该研究回顾性分析59例HeartMate 3左心室辅助装置植入者,在术后1、6、12月连续胸片测量泵角度、泵-脊柱距离和泵-膈肌深度,评估其变化及与复合不良事件的关系。结果显示较大泵角度与复合结局相关,泵深度移动呈弱关联。
要点
- 术后1-6月泵-脊柱距离和泵-膈肌深度显著变化,6-12月无显著变化。
- 1月时较大泵角度与复合不良事件相关(66.2° vs 59.0°,P=0.04)。
- 泵深度变化与复合结局呈弱关联(HR 1.019-1.020)。
核心结论
- 较大泵角度与位置相关不良事件的复合结局相关。
- 泵深度移动与复合结局呈弱关联。
主要发现
- 1至6月间,泵-脊柱距离绝对变化10.00 mm(P < 0.01),泵-膈肌深度绝对变化18.80 mm(P < 0.01)。
- 6至12月间上述参数无显著变化,泵角度各时间段均无显著变化。
- 1月泵深度和泵-脊柱距离与复合结局无关。
- 1至6月泵深度变化(HR 1.019;95%CI 1.000-1.039)和6至12月变化(HR 1.020;95%CI 1.000-1.040)与复合结局弱关联。
临床怎么用
- 连续胸片监测HeartMate 3泵位置变化,关注较大泵角度及泵深度移动,以识别位置相关不良事件风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: This study analyses the position of the HeartMate 3 left ventricular assist device on serial radiographs to assess positional change and possible correlation with adverse events. METHODS: We retrospectively analysed 59 left ventricular assist device recipients who had serial chest radiographs at 1 month, 6 months and 12 months post-implantation between November 2014 and June 2018. We measured pump angle, pump-spine distance and pump-diaphragm depth and investigated their relationship to a composite outcome of heart failure readmission, low flow alarms, stroke or inflow/outflow occlusion requiring surgical repositioning through recurrent event survival modelling. RESULTS: Between 1 and 6 months, the absolute pump-spine distance changed by 10.00 mm (P < 0.01) and the absolute pump-diaphragm depth changed by 18.80 mm (P < 0.01). These parameters did not change significantly between 6 and 12 months post-implantation. Pump angle did not change significantly over any period. Twenty-six patients experienced the composite outcome; in these patients, the median 1-month pump angle was 66.2° (interquartile range 54.5-78.0) as compared to 59.0° (interquartile range 47.0-65.0) in the 33 patients who did not have adverse events (P = 0.04). Pump depth and pump-spine distance at 1 month were not associated with the composite outcome. Change in pump depth between 1 and 6 months [hazard ratio (HR) 1.019; 95% confidence interval (CI) 1.000-1.039] and between 6 and 12 months (HR 1.020; 95% CI 1.000-1.040) were weakly associated with the composite outcome. CONCLUSIONS: Larger pump angles are associated with the composite outcome of position-related adverse events. Pump depth movement is weakly associated with the composite outcome.
引用
- Shih H, Butler C, Melehy A, et al. Serial assessment of HeartMate 3 pump position and inflow angle and effects on adverse events. Eur J Cardiothorac Surg. 2021. PMID: 33523232. doi:10.1093/ejcts/ezaa475
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。