离心式左心室辅助装置血栓事件的组织学特征
本研究对59例取出的HeartWare左心室辅助装置(HVAD)泵内血栓进行组织学分类,依据预设标准分为推测摄入性、推测新发性和不可确定性。结果显示130个组织学切片中73%为推测摄入性,27%为不可确定性,未见明确新发血栓;较大材料(>800 mm3)多见于流入道区域,较小材料(<150 mm3)常见于泵内较小流道。结论提示HVAD泵内血栓多为摄入性材料而非泵内新发血栓。
要点
- HVAD泵内血栓组织学以推测摄入性为主(73%),新发血栓未见明确证据。
- 较大血栓材料(>800 mm3)多位于流入道区域。
- 较小血栓材料(<150 mm3)常见于泵内较小流道。
核心结论
- HVAD泵内血栓常见为摄入性材料而非泵内新发血栓形成。
研究设计与样本
- 对59例取出的HVAD泵进行组织学回顾性分析,依据预设标准将血栓分类为推测摄入性、推测新发性或不可确定性。
主要发现
- 130个组织学切片中73%(95个)为推测摄入性,27%(35个)为不可确定性,无一例明确为推测新发性。
- 不可确定性的常见原因为材料不足以分析。
- 较大材料(>800 mm3)倾向于位于流入道区域。
- 最常见发现为泵内较小材料(<150 mm3),占64%(38/59例HVAD),摄入时多位于泵内较小流道。
临床怎么用
- 提示临床关注HVAD泵内血栓可能源于摄入性材料,需进一步研究摄入来源以减轻该并发症。
局限性
- 部分材料因不足无法分析,导致27%分类为不可确定性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Histology of thrombosis events in left ventricular assist devices (LVADs) may point to differences between the etiology of either ingested or de novo thrombus formation within LVADs. Materials ingested by the pump would have features suggestive of lifting and folding, whereas thrombi formed de novo would have uniform, parallel layers. This study tested this hypothesis in a cohort of explanted HeartWare Ventricular assist devices (HVADs) (Medtronic, Miami Lakes, Florida). METHODS: Histology of thrombi from 59 explanted HVAD pumps were classified as presumed ingested, presumed de novo, or undeterminable on the basis of pre-defined criteria. The apparent size and location of the thrombotic materials were noted. RESULTS: Histologically, all thrombotic materials were either presumed to be ingested (73%; 95 of 130 total histology cassettes examined) or of undeterminable origin (27%; 35 of 130 histology cassettes). Undetermined origin commonly was due to a lack of sufficient material for analysis. The larger materials (>800 mm3) tended to be in the inflow region. The most common finding was smaller thrombotic materials (<150 mm3) within the pump (64%; 38 of 59 HVADs); when these smaller materials were ingested by the pump, they were most often found within the smaller flow pathways within the pump. CONCLUSIONS: Our study suggests that the thrombi within HVAD pumps are commonly ingested materials rather than de novo thrombus formation within the pump. Further research to understand the source of this ingested material and the consideration to mitigate this complication should be considered.
引用
- Jessen SL, Kaulfus CN, Chorpenning K, et al. Histologic features of thrombosis events with a centrifugal left ventricular assist device. J Heart Lung Transplant. 2020. PMID: 33339557. doi:10.1016/j.healun.2020.10.007
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。