感染与未感染心室辅助装置导线生物膜及微间隙特征分析
研究对澳大利亚2016-2018年心脏移植患者的9条感染和13条未感染VAD导线进行分段SEM和活菌计数,发现感染导线绒面各深度均有微生物且与出口部位拭子培养一致,生物膜以微菌落形式存在于隧道内并以人组织为基质、抗微生物治疗耐药,微CT和SEM显示组织整合不足伴微间隙,体外实验观察到播散-播种机制介导的生物膜迁移。
要点
- 感染导线绒面各深度均可分离出微生物,与临床出口拭子培养一致
- 生物膜位于导线隧道内微菌落中,以人组织为基质且抗微生物治疗
- 组织整合不足伴微间隙,生物膜经播散-播种机制迁移
核心结论
- 感染导线绒面存在广泛抗微生物生物膜,与导线周围组织微间隙相关
- 增强绒面组织整合可能通过阻止生物膜经微间隙迁移来预防导线感染
研究设计与样本
- 获取澳大利亚2016-2018年心脏移植VAD患者的9条感染和13条未感染导线
- 每条导线切成11段1.5 cm,行SEM和活菌计数,培养鉴定微生物
- 用micro-CT和SEM评估宿主组织整合,体外间质生物膜实验模拟迁移并延时显微观察
主要发现
- 9条感染导线绒面各深度均分离出微生物,且与临床感染出口拭子培养一致
- SEM和micro-CT显示绒面全程组织整合不足并可见微间隙
- 临床生物膜表现为隧道内以人组织为基质的微菌落,抗微生物治疗耐药
- 观察到由播散-播种机制介导的生物膜迁移
临床怎么用
- 关注导线绒面组织整合情况,微间隙可能为生物膜迁移提供通道
- 对导线感染需考虑生物膜抗微生物特性,单纯抗微生物治疗可能不足
局限性
- 体外间质生物膜实验为模拟环境,不能完全代表体内过程
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Driveline infections remain a major complication of ventricular assist device (VAD) implantation. This study aimed to characterize in vivo microbial biofilms associated with driveline infections and host tissue integration of implanted drivelines. METHODS: A total of 9 infected and 13 uninfected drivelines were obtained from patients with VAD undergoing heart transplantation in Australia between 2016 and 2018. Each driveline was sectioned into 11 pieces of 1.5 cm in length, and each section was examined by scanning electron microscopy (SEM) and viable counts for microbial biofilms. Microorganisms were cultured and identified. Host tissue integration of clinical drivelines was assessed with micro-computed tomography (CT) and SEM. An in vitro interstitial biofilm assay was used to simulate biofilm migration in the driveline tunnel, and time-lapse microscopy was performed. RESULTS: Of the 9 explanted, infected drivelines, all had organisms isolated from varying depths along the velour section of the drivelines, and all were consistent with the swab culture results of the clinically infected exit site. SEM and micro-CT suggested insufficient tissue integration throughout the driveline velour, with microgaps observed. Clinical biofilms presented as microcolonies within the driveline tunnel, with human tissue as the sub-stratum, and were resistant to anti-microbial treatment. Biofilm migration mediated by a dispersal-seeding mechanism was observed. CONCLUSIONS: This study of explanted infected drivelines showed extensive anti-microbial-resistant biofilms along the velour, associated with microgaps between the driveline and the surrounding tissue. These data support the enhancement of tissue integration into the velour as a potential preventive strategy against driveline infections by preventing biofilm migration that may use microgaps as mediators.
引用
- Qu Y, McGiffin D, Hayward C, et al. Characterization of infected, explanted ventricular assist device drivelines: The role of biofilms and microgaps in the driveline tunnel. J Heart Lung Transplant. 2020. PMID: 32771438. doi:10.1016/j.healun.2020.07.015
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。