患者特异性计算模型优化左心室辅助装置植入后主动脉瓣关闭不全
本研究利用计算流体动力学(CFD)模拟,比较10例LVAD植入后新发主动脉瓣关闭不全(AI)患者与20例无AI患者的主动脉三维模型,发现AI组主动脉根部壁剪切应力更高,且 outflow graft 距主动脉根部更近,提示 outflow graft 位置与AI发生相关。
要点
- LVAD outflow graft距主动脉根部较近者更易发生新发AI。
- 新发AI患者主动脉根部壁剪切应力显著增高。
- outflow graft方位角与极角组合可能影响AI发生。
核心结论
- CFD模拟显示新发AI患者主动脉根部壁剪切应力更高,且outflow graft距主动脉根部更近。
研究设计与样本
- 回顾性病例对照研究,纳入2009至2018年LVAD植入患者,10例新发主动脉瓣关闭不全,20例无主动脉瓣关闭不全,基于CT构建主动脉三维模型并进行CFD模拟。
主要发现
- AI组outflow graft距主动脉根部距离为3.39 ± 0.7 cm,无AI组为4.07 ± 0.77 cm(P = .04)。
- AI组主动脉根部壁剪切应力为4.60 ± 5.70 dyne/cm2,无AI组为2.37 ± 1.20 dyne/cm2(P < .001)。
- 两组主动脉根部压力无显著差异。
- 方位角与极角单独比较相似,但高方位角与低极角组合者无AI发生。
- 定性观察AI组扰动、高壁剪切应力区域及涡流更多。
临床怎么用
- LVAD植入时可通过患者特异性CFD模拟评估outflow graft位置对主动脉根部壁剪切应力的影响,以优化配置减少新发AI风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Aortic incompetence (AI) is observed to be accelerated in the continuous-flow left ventricular assist device (LVAD) population and is related to increased mortality. Using computational fluid dynamics (CFD), we investigated the hemodynamic conditions related to the orientation of the LVAD outflow in these patients. METHOD: We identified 10 patients with new aortic regurgitation, and 20 who did not, after LVAD implantation between 2009 and 2018. Three-dimensional models of patients' aortas were created from their computed tomography scans. The geometry of the LVAD outflow graft in relation to the aorta was quantified using azimuth angles (AA), polar angles (PAs), and distance from aortic root. The models were used to run CFD simulations, which calculated the pressures and wall shear stress (rWSS) exerted on the aortic root. RESULTS: The AA and PA were found to be similar. However, for combinations of high values of AA and low values of PA, there were no patients with AI. The distance from aortic root to the outflow graft was also smaller in patients who developed AI (3.39 ± 0.7 vs 4.07 ± 0.77 cm, P = .04). There was no significant difference in aortic root pressures in the 2 groups. The rWSS was greater in AI patients (4.60 ± 5.70 vs 2.37 ± 1.20 dyne/cm2, P < .001). Qualitatively, we observed a trend of greater perturbations, regions of high rWSS, and flow eddies in the AI group. CONCLUSIONS: Using CFD simulations, we demonstrated that patients who developed de novo AI have greater rWSS at the aortic root, and their outflow grafts were placed closer to the aortic roots than those patients without de novo AI.
引用
- Kasinpila P, Kong S, Fong R, et al. Use of patient-specific computational models for optimization of aortic insufficiency after implantation of left ventricular assist device. J Thorac Cardiovasc Surg. 2020. PMID: 32653292. doi:10.1016/j.jtcvs.2020.04.164