儿童二尖瓣置换的假体选择:机械瓣与牛颈静脉带瓣管道较传统生物瓣更耐久
该单中心26年回顾性研究纳入190例20岁以下接受二尖瓣置换的患儿,共290次置换。5年和10年免于再置换率分别为76%和44%,再置换时间与假体类型显著相关,猪瓣和心包瓣风险最高。结论认为机械瓣和带支架牛颈静脉瓣较固定直径生物瓣耐久性更好。
要点
- 儿童二尖瓣置换后10年免于再置换率仅44%,再置换风险高。
- 猪瓣和心包瓣的再置换风险最高,机械瓣和牛颈静脉瓣耐久性更优。
- 较小假体尺寸和左心室发育不良也是假体失败的危险因素。
核心结论
- 儿童二尖瓣置换中,机械瓣和带支架牛颈静脉瓣较固定直径生物瓣耐久性更好。
研究设计与样本
- 单中心回顾性26年综述,纳入1992至2018年接受二尖瓣置换的20岁以下患者190例,共290次置换。
主要发现
- 中位年龄4.2岁,共行290次二尖瓣置换:机械瓣180枚、猪瓣63枚、心包瓣13枚、带支架牛颈静脉瓣34枚。
- 100枚瓣膜发生再置换,5年和10年免于再置换率分别为76%和44%。
- 再置换时间与假体类型显著相关(P < .001),猪瓣和心包瓣风险最高。
- 共9例移植和44例死亡,5年和10年无移植生存率分别为81%和76%。
- 假体类型仅在单因素分析中与死亡/移植时间显著相关(P = .021),猪瓣风险高于机械瓣。
关键数值与推荐
- 假体失败的其他危险因素包括较小假体尺寸和左心室发育不良。
- 死亡/移植的独立危险因素包括较大的指数化几何开口面积和较长的体外循环时间。
临床怎么用
- 儿童二尖瓣置换选择假体时,应优先考虑机械瓣或带支架牛颈静脉瓣,避免固定直径生物瓣。
- 对使用较小尺寸假体或合并左心室发育不良的患儿,需加强再置换风险的监测。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To determine risk factors for re-replacement and death or transplant following mitral valve replacement (MVR) in children METHODS: This is a retrospective 26-year review of patients younger than 20 years of age undergoing MVR between 1992 and 2018 at single institution. Outcomes included freedom from re-MVR and transplant-free survival. Cox proportional hazards regression models assessed association between outcomes and potential risk factors. RESULTS: At median age 4.2 years, 190 children underwent 290 MVR: 180 mechanical, 63 porcine, 13 pericardial, and 34 stented bovine jugular vein valves. Re-MVR occurred in 100 valves. Freedom from re-MVR at 5 and 10 years was 76% and 44%. Times to re-MVR were associated with prosthesis type (P < .001), with porcine and pericardial valves at greatest risk. Other risk factors for prosthetic failure included smaller prosthesis size and left ventricular hypoplasia. There were 9 transplants and 44 deaths. Transplant-free survival at 5 and 10 years was 81% and 76%. Prosthesis type was significantly associated with time to death/transplant in univariate analysis only (P = .021), with porcine at greater risk than mechanical. Independent risk factors for death/transplant included larger indexed geometric orifice area and longer bypass time. CONCLUSIONS: In pediatric patients undergoing MVR, mechanical and stented bovine jugular vein valves were associated with increased durability compared with fixed-diameter bioprosthetic alternatives.
引用
- Choi PS, Sleeper LA, Lu M, et al. Revisiting prosthesis choice in mitral valve replacement in children: Durable alternatives to traditional bioprostheses. J Thorac Cardiovasc Surg. 2020. PMID: 32713632. doi:10.1016/j.jtcvs.2020.04.173
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。