可回收覆膜金属分段Y型气道支架治疗隆突或主支气管胃呼吸道瘘
本研究评估3D打印定制的可回收覆膜金属分段Y型气道支架治疗食管切除术后隆突或主支气管胃呼吸道瘘的可行性。支架留置中位时间64天,取出后瘘口均愈合,无支架断裂。
要点
- 3D打印定制可回收覆膜金属分段Y型支架可用于隆突或主支气管胃呼吸道瘘。
- 6例患者植入7枚支架均成功置入和取出,瘘口完全封闭,吸入性肺炎均控制。
- 支架留置中位时间64天,取出后瘘口均愈合,无支架断裂。
核心结论
- 3D打印定制的可回收覆膜金属分段Y型气道支架治疗隆突或主支气管胃呼吸道瘘看来可行。
研究设计与样本
- 研究设计为可行性评估,纳入6例食管切除术后胃呼吸道瘘合并吸入性肺炎患者,均植入个体化3D打印定制的可回收覆膜金属分段Y型气道支架。
主要发现
- 6例患者共7枚Y型支架成功植入并取出,所有支架膨胀良好,瘘口完全封闭。
- 6例患者吸入性肺炎均得到控制。
- 支架留置时间为64天(四分位距52-69天)。
- 取出后支气管镜、胃镜和胸部CT显示所有患者瘘口均愈合,无支架断裂。
关键数值与推荐
- 痰潴留是支架置入后最常见的并发症,发生率为33.33%,经支气管镜下吸引治疗。
- 1例患者(16.7%)发现肉芽组织过度增生,经冷冻治疗。
临床怎么用
- 对于食管切除术后隆突或主支气管胃呼吸道瘘患者,可考虑使用3D打印定制的可回收覆膜金属分段Y型气道支架封堵瘘口并控制吸入性肺炎。
- 支架置入后需警惕痰潴留和肉芽组织增生,及时行支气管镜下吸引或冷冻治疗。
- 瘘口愈合或出现支架相关并发症时可取出支架,取出后需复查支气管镜、胃镜和胸部CT确认瘘口愈合。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To evaluate the feasibility of new retrievable covered metallic segmented Y airway stents modified with 3-dimensional (3D) printing for gastrorespiratory fistula involving carina or main bronchi. METHODS: We designed a new retrievable covered metallic segmented Y airway stent to fit the anatomical characteristics of the carina region in individual patients. All stents were individually customized based on a 3D-printed mold. Six patients with gastrorespiratory fistula and aspiration pneumonia after esophagectomy underwent the stent implantation. The stents were retrieved when the fistula was cured or stent-related complications occurred. RESULTS: Seven Y stents were successfully implanted and removed in 6 patients. All stents expanded well, and the fistulas were completely sealed. Aspiration pneumonia was controlled in 6 patients. The median Karnofsky Performance Status scores significantly improved after stenting compared with those before stent implantation (P = .024). Sputum retention was the most common complication after stenting and was treated with aspiration under bronchoscopy (33.33%). Excessive granulation tissue proliferation was found in 1 patient (16.7%) and was treated with cryotherapy. The indwelling time of the stent was 64 days (interquartile range, 52-69 days). After stent removal, bronchoscopy, gastroscopy, and computed tomography of the chest showed cured fistulas in all patients, and no stents showed fractures. CONCLUSIONS: Retrievable covered metallic segmented Y airway stents modified with 3D printing appear to be feasible for the treatment of gastrorespiratory fistula involving carina or main bronchi.
引用
- Huang W, Shan Q, Wu Z, et al. Retrievable covered metallic segmented Y airway stent for gastrorespiratory fistula of carina or main bronchi. J Thorac Cardiovasc Surg. 2020. PMID: 32340805. doi:10.1016/j.jtcvs.2020.03.019
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。