经导管主动脉瓣-in-经导管主动脉瓣的窦部血流动力学研究
本研究在脉冲复制器中评估6种经导管主动脉瓣-in-经导管主动脉瓣(TAV-in-TAV)配置对窦部血流动力学和冲刷的影响。结果显示TAV-in-TAV高度依赖原植入瓣膜和新植入瓣膜的组合,窦部冲刷与瓣中瓣及TAV置换相比未见显著恶化。
要点
- TAV-in-TAV的血流动力学高度依赖原植入瓣膜与新植入瓣膜的组合。
- 窦部冲刷在TAV-in-TAV后未见显著恶化。
- 需进一步研究优化瓣膜尺寸和选择。
核心结论
- TAV-in-TAV的血流动力学高度依赖原植入瓣膜和新植入瓣膜的组合。
- 与瓣中瓣和TAV置换相比,TAV-in-TAV后窦部冲刷未见显著恶化。
研究设计与样本
- 在脉冲复制器中选择6种TAV-in-TAV配置进行评估,并进行粒子图像测速。
主要发现
- 收缩期最高流速见于SAPIEN-in-SAPIEN(0.7 m/s),最低见于Evolut 26-in-Evolut 26(0.2 m/s)。
- 瓣叶附近最高剪切应力见于Evolut 23-in-SAPIEN(1.45 Pa),最低见于Evolut 26-in-Evolut 26(0.55 Pa)。
- 所有病例各窦的冲刷几乎相同(<2.5个周期)。
临床怎么用
- TAV-in-TAV方案选择需考虑原植入瓣膜与新植入瓣膜的配对对窦部血流动力学的潜在影响。
局限性
- 需进一步研究优化瓣膜尺寸和选择。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: This study evaluated the effect of transcatheter aortic valve (TAV)-in-TAV on sinus hemodynamics and washout. With TAV becoming the standard procedure for aortic valve replacement and with the limited valve durability, a second intervention is necessary (TAV-in-TAV) after first TAV failure. METHODS: Six arrangements of TAV-in-TAV were chosen for this study as follows: (1) Evolut 23 (Medtronic, Minneapolis, MN) in Evolut 26, (2) Evolut 23 in SAPIEN 3 23 (Edwards Lifesciences, Irvine, CA), (3) Evolut 26 in Evolut 26, (4) Evolut 26 in SAPIEN 23, (5) SAPIEN 3 23 in Evolut 26, and (6) SAPIEN 3 23 in SAPIEN 3 23. These TAV-in-TAV configurations were assessed in a pulse duplicator. Particle image velocimetry was performed. RESULTS: During systole, (1) the highest velocity was found with SAPIEN-in-SAPIEN (0.7 m/s) and the lowest was with Evolut 26-in-Evolut 26 (0.2 m/s); (2) the highest shear stress magnitude near the leaflet was with Evolut 23-in-SAPIEN (1.45 Pa) and the lowest was with Evolut 26-in-Evolut 26 (0.55 Pa); and (3) washout was almost equal in all sinuses of these cases (<2.5 cycles). CONCLUSIONS: This study shows that TAV-in-TAV is highly dependent on the valve that is originally implanted and the valve to be implanted. Washout is not significantly degraded after TAV-in-TAV compared with valve-in-valve and TAV replacement. Further studies are needed to optimize valve size and selection.
引用
- Hatoum H, Lilly S, Maureira P, et al. Sinus Hemodynamics After Transcatheter Aortic Valve in Transcatheter Aortic Valve. Ann Thorac Surg. 2020. PMID: 32179042. doi:10.1016/j.athoracsur.2020.02.016
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。