腔肺辅助装置:长期逆转Fontan悖论
本研究设计了一种基于von Karman粘性泵的旋转血泵,植入全腔肺连接(TCPC)内,旨在同时降低体静脉压力并提高肺动脉压力,从而改善前负荷和心输出量。体外模拟循环测试显示,该泵在所有四个TCPC方向均能增加流量,压力升高6至8 mm Hg,最高转速5600 rpm时达14 mm Hg,停转时无压力损失。该技术有望长期维持单心室Fontan循环的双心室生理,适用于右心循环效率低下为主且心室收缩功能保留的Fontan失败患者。
要点
- 植入式腔肺辅助泵可同时降低体静脉压并升高肺动脉压。
- 体外测试中压力升高6-8 mm Hg,最高转速下达14 mm Hg。
- 停转时无压力损失,可默认作为被动血流分流器。
核心结论
- 该血泵技术可逆转Fontan悖论,可能实现单心室Fontan循环的长期双心室维持。
研究设计与样本
- 基于von Karman粘性泵设计旋转血泵,植入全腔肺连接(TCPC),在体外Fontan模拟循环回路中用血液类似物进行压力-流量性能表征。
主要发现
- 泵在全部规定操作范围内工作,在TCPC所有4个方向均增加流量。
- 压力升高6至8 mm Hg可轻松实现,最高转速5600 rpm时达14 mm Hg。
- 性能在广泛心输出量范围内一致。
- 停转状态(0 rpm)下TCPC无明显压力损失。
临床怎么用
- 适用于右心循环效率低下为主且心室收缩功能保留的Fontan失败患者。
- 也可在临床Fontan失败前作为健康维护,预防Fontan疾病进展。
局限性
- 研究为体外模拟循环测试,未提供临床植入数据。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Fontan circulatory inefficiency can be addressed by replacing the missing subpulmonary power source to reverse the Fontan paradox. An implantable cavopulmonary assist device is described that will simultaneously reduce systemic venous pressure and increase pulmonary arterial pressure, improving preload and cardiac output, in a univentricular Fontan circulation on a long-term basis. METHODS: A rotary blood pump that was based on the von Karman viscous pump was designed for implantation into the total cavopulmonary connection (TCPC). It will impart modest pressure energy to augment Fontan flow without risk of obstruction. In the event of rotational failure, it is designed to default to a passive flow diverter. Pressure-flow performance was characterized in vitro in a Fontan mock circulatory loop with blood analog. RESULTS: The pump performed through the fully specified operating range, augmenting flow in all 4 directions of the TCPC. Pressure rise of 6 to 8 mm Hg was readily achieved, ranging to 14 mm Hg at highest speed (5600 rpm). Performance was consistent across a wide range of cardiac outputs. In stalled condition (0 rpm), there was no discernible pressure loss across the TCPC. CONCLUSIONS: A blood pump technology is described that can reverse the Fontan paradox and may permit a surgical strategy of long-term biventricular maintenance of a univentricular Fontan circulation. The technology is intended for Fontan failure in which right-sided circulatory inefficiencies predominate and ventricular systolic function is preserved. It may also apply before clinical Fontan failure as health maintenance to preempt the progression of Fontan disease.
引用
- Rodefeld MD, Marsden A, Figliola R, et al. Cavopulmonary assist: Long-term reversal of the Fontan paradox. J Thorac Cardiovasc Surg. 2019. PMID: 31564543. doi:10.1016/j.jtcvs.2019.06.112
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。