文献速递瓣膜外科先天性心脏病外科

牛心包瓣膜在肺动脉位与主动脉位的压力行为对比

Pulmonary versus aortic pressure behavior of a bovine pericardial valve.
PubMed 文献 · J Thorac Cardiovasc Surg · 2019-07-05

本研究用脉冲复制器在肺动脉压力条件(15/5、28/11、73/32 mm Hg)与正常主动脉压力(120/80 mm Hg)下评估21 mm和25 mm Carpentier Edwards Perimount Magna Ease瓣膜的流体动力学特性。结果显示正常肺动脉条件下21 mm瓣膜漏孔面积为0.020±0.012 mm2、25 mm瓣膜为0.054±0.041 mm2(P=.03),肺动脉低血压及正常肺动脉条件下瓣膜关闭容积显著低于正常主动脉条件(P<.05)。作者认为正常肺动脉压力下该瓣膜的流体动力学特性与主动脉条件显著不同,需进一步研究其与人工瓣膜衰败的关系。

要点

  • 正常肺动脉压力下Magna Ease瓣膜的流体动力学特性与主动脉条件显著不同。
  • 21 mm与25 mm瓣膜在正常肺动脉条件下的漏孔面积存在显著差异(P=.03)。
  • 肺动脉低血压及正常肺动脉条件下瓣膜关闭容积显著低于正常主动脉条件(P<.05)。

核心结论

  • 正常肺动脉压力条件下Magna Ease瓣膜的流体动力学特性与主动脉条件显著不同,尚需进一步研究这些结果是否与人工瓣膜衰败相关。

研究设计与样本

  • 使用脉冲复制器评估21 mm和25 mm Magna Ease瓣膜的流体动力学特性。
  • 在肺动脉压力条件15/5 mm Hg、28/11 mm Hg、73/32 mm Hg及正常主动脉压力120/80 mm Hg参考条件下测试。
  • 用Phantom MIRO M320S高速相机以600帧/秒获取动态图像并分析闭合状态瓣膜面积。

主要发现

  • 正常肺动脉条件(28/11 mm Hg)下漏孔面积21 mm瓣膜为0.020±0.012 mm2,25 mm瓣膜为0.054±0.041 mm2(P=.03)。
  • 瓣膜流体动力学特性在肺动脉与主动脉测试条件之间存在差异。
  • 肺动脉低血压和正常肺动脉条件下瓣膜关闭容积显著低于正常主动脉条件(P<.05)。

临床怎么用

  • 提示该瓣膜超说明书用于肺动脉位时,其流体动力学行为与主动脉位不同,临床随访需关注瓣膜功能。

局限性

  • 研究为脉冲复制器体外实验,非体内临床数据。
  • 样本仅涉及21 mm和25 mm两种规格的Magna Ease瓣膜。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The Carpentier Edwards Perimount Magna Ease aortic valvular prosthesis (Edwards Lifesciences, Irvine, Calif) has been among the most frequently and successfully used tissue prosthetic cardiac valves. Furthermore, this prosthesis has been used off-label in the pulmonary position. Until now, there has been a paucity of data regarding the functioning of tissue prosthetic valves under pulmonary conditions. METHODS: Using a pulse duplicator, hydrodynamic characteristics of a 21-mm and 25-mm Magna Ease valve were evaluated. Among parameters evaluated were leakage orifice area, closing time (ie, time required to close), and leakage duration. This procedure was performed under different pulmonic pressure conditions (15/5 mm Hg, 28/11 mm Hg, 73/32 mm Hg) and normal aortic pressure (120/80 mm Hg) as a reference. Moving images were obtained using a Phantom MIRO M320S high-speed camera (Vision Research Inc, Wayne, NJ) at 600 frames per second and used to analyze valve area in closed position. RESULTS: Under normal pulmonic conditions (28/11 mm Hg) the leakage orifice area was 0.020 ± 0.012 mm2 for the 21-mm valve and 0.054 ± 0.041 mm2 for the 25-mm valve (P = .03). Hydrodynamic characteristics of the valves differed between pulmonary and aortic testing condition. Valve closing volumes were significantly lower under pulmonary hypotension and normal pulmonary conditions than under normal aortic conditions (P < .05). CONCLUSIONS: Under normal pulmonary pressure conditions, the hydrodynamic characteristics of Magna Ease valves are significantly different compared with aortic conditions. Further research is needed to determine whether these results are associated with prosthetic valve failure.

来源:J Thorac Cardiovasc Surg,查看原文。版权归原作者所有。

引用

  1. Pragt H, van Melle JP, Verkerke GJ, et al. Pulmonary versus aortic pressure behavior of a bovine pericardial valve. J Thorac Cardiovasc Surg. 2019. PMID: 31383558. doi:10.1016/j.jtcvs.2019.05.084

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。