文献速递瓣膜外科

早期第一代Trifecta瓣膜衰败:病例系列与文献综述

Early First-Generation Trifecta Valve Failure: A Case Series and a Review of the Literature.
PubMed 文献 · Ann Thorac Surg · 2019-07-20

该研究回顾2012至2014年107例Trifecta生物瓣植入,发现7例(6.5%)早期衰败,平均耐久时间51±16个月,表现为瓣膜狭窄或严重反流,病理为环周血管翳形成伴无冠瓣撕裂和瓣叶钙化。早期衰败发生率48个月为3.1%,72个月为13.1%。终末期肾病和患者-人工瓣膜不匹配者衰败率更高,部分瓣叶撕裂患者需紧急手术。

要点

  • Trifecta瓣膜早期衰败表现为狭窄或严重反流,病理为环周血管翳、瓣叶撕裂和钙化。
  • 终末期肾病和患者-人工瓣膜不匹配者应密切随访。
  • 部分瓣叶撕裂患者可能需紧急再次手术。

核心结论

  • 早期Trifecta瓣膜衰败由瓣叶撕裂或钙化导致,终末期肾病和不匹配者需密切随访,部分瓣叶撕裂者需紧急手术。

研究设计与样本

  • 回顾性病例系列,2012至2014年共植入107例Trifecta生物瓣,发生7例衰败(6.5%)。

主要发现

  • 7例衰败中19mm瓣1例、21mm瓣5例、23mm瓣1例,平均耐久时间51±16个月。
  • 首次超声平均有效瓣口面积指数0.96±0.26,表现为瓣膜狭窄3例、严重反流4例。
  • 6例再次手术,常见病理为环周血管翳形成伴无冠瓣撕裂和瓣叶钙化。
  • 衰败患者术前终末期肾病和术后患者-人工瓣膜不匹配比例更高。
  • 早期衰败发生率48个月为3.1%,72个月为13.1%。

临床怎么用

  • 对植入Trifecta瓣膜且合并终末期肾病或患者-人工瓣膜不匹配者应加强超声随访。
  • 发现瓣叶撕裂相关衰败时应警惕需紧急手术的可能。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The Trifecta valve (Abbott, St Paul, MN) has excellent hemodynamic performance with acceptable rates of freedom from structural valve degeneration. However, some recent studies have reported early Trifecta valve failure. Here, we report a case series of seven Trifecta valve failures with a review of the literature. METHODS: Of 107 implantations of Trifecta bioprostheses between 2012 and 2014, we encountered seven Trifecta valve failures (6.5%). Failure of a 19-mm Trifecta valve occurred in 1 patient, failure of a 21-mm Trifecta valve occurred in 5 patients, and failure of a 23-mm Trifecta valve occurred in 1 patient. The mean duration of valve durability was 51 ± 16 months. The mean effective orifice area index on the first echocardiogram after Trifecta valve implantation was 0.96 ± 0.26. The mode of presentation was prosthetic valve stenosis in 3 patients and severe aortic regurgitation in 4 patients. RESULTS: Six patients underwent redo surgical aortic valve replacement. The common pathologic findings were circumferential pannus formation with noncoronary cusp tear and leaflet calcification. The rates of preoperative end-renal stage disease and postoperative prosthesis-patient mismatch were higher in patients with Trifecta valve failure. The incidence of early Trifecta valve failure was 3.1% at 48 months and 13.1% at 72 months. CONCLUSIONS: In our experience, early Trifecta valve failure was caused by cusp tears or leaflet calcification. Patients with end-renal stage disease and prosthesis-patient mismatch should be closely followed. Some patients with cusp tears may require urgent surgery.

来源:Ann Thorac Surg,查看原文。版权归原作者所有。

引用

  1. Kaneyuki D, Nakajima H, Asakura T, et al. Early First-Generation Trifecta Valve Failure: A Case Series and a Review of the Literature. Ann Thorac Surg. 2019. PMID: 31336064. doi:10.1016/j.athoracsur.2019.05.073

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。